A Deep Neural Network for Interpreting Wearable Electrocardiogram Data in Atrial Fibrillation: Prospective Observational Diagnostic Accuracy Study
JMIR mHealth and uHealth · 23.04.2026 · 14 · e82475-e82475
DOI: 10.2196/82475 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift JMIR mHealth and uHealth, online erschienen am 23.04.2026. Verfasst von Olli A Rantula und 11 weiteren Autor:innen. Beteiligte Einrichtung: School of Medicine, Faculty of Health Sciences, University of Eastern Finland, Yliopistonranta 1, PO BOX 1627, Kuopio, 70211, Finland, 358 0294451111. Hauptthemen laut MEDLINE: Atrial Fibrillation, Electrocardiography, Wearable Electronic Devices, Neural Networks, Computer. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
Atrial fibrillation (AF) and atrial flutter (AFL) are common arrhythmias associated with the risk of ischemic stroke, which can be reduced with anticoagulation therapy. Thus, early diagnosis of AF and AFL is essential. However, diagnosis may be challenging due to the paroxysmal and asymptomatic nature of these arrhythmias.
Objective
Current diagnostic workflows involve time-consuming and resource-intensive manual review of noisy signals and prolonged recordings. We evaluated a mobile system that combines a wireless wearable single-lead chest strap electrocardiogram (ECG) and a novel deep neural network (DNN)–based artificial intelligence (AI) method for detecting AF/AFL episodes, AF/AFL burden, and rhythm change and estimated the delay in the detection of rhythm change from AF/AFL to sinus rhythm. We also assessed the rhythm classification performance.
Methods
A total of 116 patients with recent-onset AF or AFL undergoing cardioversion were monitored using a mobile single-lead chest strap ECG system. Simultaneously, a 3-lead Holter ECG served as the reference. The DNN-based AI analyzed the single-lead chest strap ECG data to detect AF/AFL, non-AF/AFLrhythm, and noninterpretable segments, as well as to estimate AF/AFL burden and detect rhythm change. Performance metrics included sensitivity, specificity, positive predictive value, negative predictive value, and intraclass correlation coefficient for AF and AFL burden estimation.
Results
The sensitivity and specificity for detecting AF/AFL were 91.9% (204.9/223.0 h) and 99.6% (242.4/243.5 h), respectively. The sensitivity for detecting AF was 96.2% (191.5/199.0 h), whereas it was 55.8% (13.4/24.0 h) for detecting AFL. The positive predictive value and negative predictive value for AF/AFL detection were 99.5% (204.9/206.0 h) and 93.1% (242.4/260.5 h), respectively. The intraclass correlation coefficient between the AF/AFL burden estimated by the DNN-based AI method and that derived from the physician-interpreted reference ECG was 0.96 (95% CI: 0.94‐0.97; P
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Medicine, Faculty of Health Sciences, University of Eastern Finland, Yliopistonranta 1, PO BOX 1627, Kuopio, 70211, Finland, 358 0294451111.
- Doctoral School, Faculty of Health Sciences, University of Eastern Finland, Kuopio, Finland.
- Heart Center, Kuopio University Hospital, Kuopio, Finland.
- Department of Technical Physics, Faculty of Science and Forestry, University of Eastern Finland, Kuopio, Finland.
- Department of Anaesthesiology and Intensive Care, North Karelia Central Hospital, Joensuu, Finland.
- Heart Center, North Karelia Central Hospital, Joensuu, Finland.
ORCID-Profile: Olli A Rantula, Jukka A Lipponen, Jari Halonen, Helena Jäntti, Tuomas T Rissanen, Mika P Tarvainen, Noora S Naukkarinen, Eemu-Samuli Seljola, Onni E Santala, Jagdeep Sedha, Tero J Martikainen, Juha E K Hartikainen
Zitieren
APA
Rantula, O. A., Lipponen, J. A., Halonen, J., Jäntti, H., Rissanen, T. T., Tarvainen, M. P., Naukkarinen, N. S., Seljola, E. S., Santala, O. E., Sedha, J., Martikainen, T. J., & Hartikainen, J. E. K. (2026). A Deep Neural Network for Interpreting Wearable Electrocardiogram Data in Atrial Fibrillation: Prospective Observational Diagnostic Accuracy Study. JMIR mHealth and uHealth, 14, e82475-e82475. https://doi.org/10.2196/82475
Vancouver
Rantula OA, Lipponen JA, Halonen J, Jäntti H, Rissanen TT, Tarvainen MP, et al. A Deep Neural Network for Interpreting Wearable Electrocardiogram Data in Atrial Fibrillation: Prospective Observational Diagnostic Accuracy Study. JMIR mHealth and uHealth. 2026;14:e82475-e82475. doi:10.2196/82475
BibTeX
@article{Rantula2026,
author = {Rantula, Olli A and Lipponen, Jukka A and Halonen, Jari and Jäntti, Helena and Rissanen, Tuomas T and Tarvainen, Mika P and Naukkarinen, Noora S and Seljola, Eemu-Samuli and Santala, Onni E and Sedha, Jagdeep and Martikainen, Tero J and Hartikainen, Juha E K},
title = {A Deep Neural Network for Interpreting Wearable Electrocardiogram Data in Atrial Fibrillation: Prospective Observational Diagnostic Accuracy Study},
journal = {JMIR mHealth and uHealth},
year = {2026},
volume = {14},
pages = {e82475--e82475},
doi = {10.2196/82475}
}
EUVIMED