Validation of Smartwatches Integrated With Photoplethysmography for Continuous Evaluation of Atrial Fibrillation Burden

Song Zuo, Le Zhou, Han Feng, Christian Massad, Michel Abou Khalil, Jinglei Wang, Jue Wang, Yize Zhao, Fadong Li, Xiangyi Kong, Chao Jiang, Xin Li und 14 weitere

JACC: Clinical Electrophysiology · 2026-09 · 12 (9) · 2226-2237

DOI: 10.1016/j.jacep.2026.03.036 ↗

Kurz gefasst

Beobachtungsstudie in der Zeitschrift JACC: Clinical Electrophysiology, 2026-09. Verfasst von Song Zuo und 25 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Beijing Anzhen Hospital, Capital Medical University, Beijing, China; National Clinical Research Center for Cardiovascular D…. Hauptthemen laut MEDLINE: Atrial Fibrillation, Photoplethysmography, Wearable Electronic Devices. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 32 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Smartwatches integrated with photoplethysmography (PPG) can effectively identify the occurrence of atrial fibrillation (AF). However, the accuracy of their continuous monitoring of AF burden is still unclear. OBJECTIVES: This study sought to verify the effectiveness of a smartwatch integrated with the PPG algorithm in continuously monitoring AF burden. METHODS: This prospective study included patients diagnosed with AF at Beijing Anzhen Hospital between January and December 2024. Each participant continuously wore a smartwatch integrated with the PPG algorithm on the day prior to radiofrequency catheter ablation, together with patch-based electrocardiography (P-ECG) as the reference to validate algorithm performance for estimation of AF burden. The watch-based PPG data were segmented into 30-second intervals and verified with the data from P-ECG to analyze its accuracy, sensitivity, and specificity. RESULTS: The study recruited 728 participants, with a mean age of 62.0 ± 10.5 years, of whom 412 (56.6%) had paroxysmal AF. The average monitoring time for each participant was 20.4 ± 4.5 hours. The overall validity rate for P-ECG was 96.2%, while the overall validity rate for watch-based PPG was 62.5%. After segmentation, a total of 1,440,826 PPG records were generated. At the interval level, sensitivity was 98.70% (95% CI: 98.66%-98.73%) and specificity was 99.56% (95% CI: 99.54%-99.57%). At the individual level, sensitivity was 91.73% (95% CI: 88.89%-94.15%) and specificity was 96.96% (95% CI: 95.09%-98.59%). The Bland-Altman analysis showed excellent agreement between the AF burden estimates from PPG and electrocardiography (ECG), with a mean difference of -1.34% and 95% limits of agreement ranging from -6.45% to 3.77%. A strong correlation (r = 0.999) was observed between PPG-based and ECG-measured AF burden. CONCLUSIONS: Smartwatches integrated with the PPG algorithm offer a practical and noninvasive approach for continuous AF burden monitoring. The results suggest that PPG can be an effective approach for long-term AF management, providing an accessible alternative to traditional ECG monitoring. (Study on the Effectiveness of AF Analysis Software in AF Rhythm Monitoring [PPG-AF]; NCT05333380).

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Atrial FibrillationPhotoplethysmographyWearable Electronic DevicesAgedFemaleHumansMaleMiddle AgedAlgorithmsElectrocardiographyProspective StudiesReproducibility of ResultsSensitivity and Specificity

Stichwörter der Autor:innen

atrial fibrillation burdenmonitoringphotoplethysmographysmartwatch

Autor:innen und Institutionen

  • Department of Cardiology, Beijing Anzhen Hospital, Capital Medical University, Beijing, China; National Clinical Research Center for Cardiovascular Diseases, Beijing, China.
  • Cardiology Department, Tulane University School of Medicine, New Orleans, Louisiana, USA.
  • Chengdu CVhealth Science and Technology Co., Ltd., Chengdu, China.
  • Cardiology Department, Tulane University School of Medicine, New Orleans, Louisiana, USA. Electronic address: nmarrouche@tulane.edu.
  • Department of Cardiology, Beijing Anzhen Hospital, Capital Medical University, Beijing, China; National Clinical Research Center for Cardiovascular Diseases, Beijing, China. Electronic address: chshma@vip.sina.com.

Zitieren

APA

Zuo, S., Zhou, L., Feng, H., Massad, C., Khalil, M. A., Wang, J., Wang, J., Zhao, Y., Li, F., Kong, X., Jiang, C., Li, X., Xia, S., He, L., Liu, X., Guo, X., Zhou, N., Li, S., Jiang, C., …, & Ma, C. (2026). Validation of Smartwatches Integrated With Photoplethysmography for Continuous Evaluation of Atrial Fibrillation Burden. JACC: Clinical Electrophysiology, 12(9), 2226-2237. https://doi.org/10.1016/j.jacep.2026.03.036

Vancouver

Zuo S, Zhou L, Feng H, Massad C, Khalil MA, Wang J, et al. Validation of Smartwatches Integrated With Photoplethysmography for Continuous Evaluation of Atrial Fibrillation Burden. JACC: Clinical Electrophysiology. 2026;12(9):2226-2237. doi:10.1016/j.jacep.2026.03.036

BibTeX
@article{Zuo2026,
  author = {Zuo, Song and Zhou, Le and Feng, Han and Massad, Christian and Khalil, Michel Abou and Wang, Jinglei and Wang, Jue and Zhao, Yize and Li, Fadong and Kong, Xiangyi and Jiang, Chao and Li, Xin and Xia, Shijun and He, Liu and Liu, Xiaoxia and Guo, Xueyuan and Zhou, Ning and Li, Songnan and Jiang, Chenxi and Tang, Ribo and Sang, Caihua and Long, Deyong and Du, Xin and Dong, Jianzeng and Marrouche, Nassir F and Ma, Changsheng},
  title = {Validation of Smartwatches Integrated With Photoplethysmography for Continuous Evaluation of Atrial Fibrillation Burden},
  journal = {JACC: Clinical Electrophysiology},
  year = {2026},
  volume = {12},
  number = {9},
  pages = {2226--2237},
  doi = {10.1016/j.jacep.2026.03.036}
}
Kostenlos starten →