Accuracy and Variability of a Commercial Markerless Motion Capture System Compared to a Pressure Mat for Weight Distribution in Standing: Cross-Sectional Observational Study
JMIR Formative Research · 2025
DOI: 10.2196/73575 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift JMIR Formative Research, 2025. Verfasst von Lisa Sheehy und 4 weiteren Autor:innen. Beteiligte Einrichtung: Bruyère Health Research Institute, Ottawa, ON, Canada. Hauptthemen laut MEDLINE: Standing Position, Weight-Bearing. Die Arbeit nennt 42 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
Commercial markerless motion capture (MMC) systems show promise for use in rehabilitation and have been validated for the assessment of various parameters. However, no prior studies have evaluated MMC systems to detect stance asymmetry.
Objective
The objective of this study was to assess the accuracy and variability of the Jintronix Weight Shift Tool MMC system to estimate the percentage of weight borne on each foot.
Methods
Twelve healthy younger adults, 12 healthy older adults, and 12 people living with stroke were recruited for this cross-sectional study. The percentage of weight borne on each foot was simultaneously recorded by the Weight Shift Tool and a validated pressure mat during 2 series (raising the arm to capture the recording and without arm raise) with left lean, right lean, and equal stance. Agreement between the Weight Shift Tool and the pressure mat was assessed using Bland-Altman analyses.
Results
Bias was greatest for older adults for all stances except for right lean without arm raise. Variability was greatest for people living with stroke for all stances except for left lean with arm raise. On average, the limits of agreement were narrower during equal stance. Although bias between the Weight Shift Tool and the pressure mat was small to moderate (0.0%-11.7%), the limits of agreement were wide (12.8%-33.6% above and below the bias).
Conclusions
The Weight Shift Tool is not clinically acceptable for the estimation of the percentage of weight on each foot due to high variability. Investigation of other MMC systems is required to confirm the validity of MMC for clinical assessment.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Bruyère Health Research Institute, Ottawa, ON, Canada.
- Interdisciplinary School of Health Sciences, Faculty of Health Sciences, University of Ottawa, Ottawa, ON, Canada.
- School of Rehabilitation Sciences, Faculty of Health Sciences, University of Ottawa, Ottawa, ON, Canada.
- Department of Systems and Computer Engineering, Faculty of Engineering and Design, Carleton University, Ottawa, ON, Canada.
- School of Human Kinetics, Faculty of Health Sciences, University of Ottawa, Ottawa, ON, Canada.
- Department of Medicine, Division of Physical Medicine and Rehabilitation, Faculty of Medicine, University of Ottawa, Ottawa, ON, Canada.
ORCID-Profile: Lisa Sheehy, Emma Gal-Dev, Heidi Sveistrup, Martin Bilodeau, Hillel Finestone
Zitieren
APA
Sheehy, L., Gal-Dev, E., Sveistrup, H., Bilodeau, M., & Finestone, H. (2025). Accuracy and Variability of a Commercial Markerless Motion Capture System Compared to a Pressure Mat for Weight Distribution in Standing: Cross-Sectional Observational Study. JMIR Formative Research. https://doi.org/10.2196/73575
Vancouver
Sheehy L, Gal-Dev E, Sveistrup H, Bilodeau M, Finestone H. Accuracy and Variability of a Commercial Markerless Motion Capture System Compared to a Pressure Mat for Weight Distribution in Standing: Cross-Sectional Observational Study. JMIR Formative Research. 2025. doi:10.2196/73575
BibTeX
@article{Sheehy2025,
author = {Sheehy, Lisa and Gal-Dev, Emma and Sveistrup, Heidi and Bilodeau, Martin and Finestone, Hillel},
title = {Accuracy and Variability of a Commercial Markerless Motion Capture System Compared to a Pressure Mat for Weight Distribution in Standing: Cross-Sectional Observational Study},
journal = {JMIR Formative Research},
year = {2025},
doi = {10.2196/73575}
}
EUVIMED