EUVIMEDObjevte EUVIMED
Open access

Effectiveness of Using a Digital Wearable Plantar Pressure Device to Detect Muscle Fatigue: Within-Subject, Repeated Measures Experimental Design

Fu-Yu Chen iD, Tzu-Yao Lin iD, Yi-Cheng Huang iD, Evina Widianawati iD

JMIR Human Factors · 2025

DOI: 10.2196/65578 ↗

At a glance

Journal article in JMIR Human Factors, 2025. Written by Fu-Yu Chen and 3 co-authors. Institution: Department of Biomedical Engineering, Chung Yuan Christian University, No. 200, Zhongbei Road, Zhongli District, Toayuan City, 32023, Taiwan, 886 325…. Major MEDLINE topics: Muscle Fatigue, Wearable Electronic Devices, Electromyography, Pressure. Cited 6 times so far (Crossref); 50 references. The full text is freely available in PubMed Central.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Background

Muscle fatigue, characterized by reduced force generation during repetitive contractions, impacts older adults doing daily activities and athletes during sports activities. While various sensors detect muscle fatigue via muscle activity, biochemical markers, and kinematic parameters, a real-time wearable solution with high usability remains limited. Plantar pressure monitoring detects muscle fatigue through foot loading changes, seamlessly integrating into footwear to improve the usability and compliance for home-based monitoring.

Objective

This study aimed to investigate the effects of muscle fatigue on plantar pressure measurements using a self-developed wearable plantar pressure system.

Methods

Twelve healthy participants completed a 5-minute calf muscle fatigue protocol. The plantar pressures and surface electromyography (sEMG) activity of the gastrocnemius muscles were recorded before and after exercise. The plantar pressures at 6 regions and the median frequency (MDF) of sEMG were analyzed to quantify fatigue.

Results

The self-developed foot pressure system showed a significant decrease in plantar pressure peak values at the heel of the left ( P =.003) and right feet ( P =.001) and at the lateral toe of the left ( P =.001) and right feet ( P =.026). A significant increase was observed at the metatarsal head of both the left foot ( P =.001) and the right foot ( P =.017). The MDF of sEMG signals significantly decreased in the left ( P =.001) and right gastrocnemius ( P

Subjects

MeSH terms (MEDLINE)

Muscle FatigueWearable Electronic DevicesElectromyographyPressureFootHumansMaleAdultFemaleMuscle, SkeletalBiomechanical Phenomena

Author keywords

home-based monitoringmuscle fatigueplantar pressure sensorswearable devices

Authors and institutions

  • Department of Biomedical Engineering, Chung Yuan Christian University, No. 200, Zhongbei Road, Zhongli District, Toayuan City, 32023, Taiwan, 886 32564507.

ORCID profiles: Fu-Yu Chen, Tzu-Yao Lin, Yi-Cheng Huang, Evina Widianawati

Cite

APA

Chen, F. Y., Lin, T. Y., Huang, Y. C., & Widianawati, E. (2025). Effectiveness of Using a Digital Wearable Plantar Pressure Device to Detect Muscle Fatigue: Within-Subject, Repeated Measures Experimental Design. JMIR Human Factors. https://doi.org/10.2196/65578

Vancouver

Chen FY, Lin TY, Huang YC, Widianawati E. Effectiveness of Using a Digital Wearable Plantar Pressure Device to Detect Muscle Fatigue: Within-Subject, Repeated Measures Experimental Design. JMIR Human Factors. 2025. doi:10.2196/65578

BibTeX
@article{Chen2025,
  author = {Chen, Fu-Yu and Lin, Tzu-Yao and Huang, Yi-Cheng and Widianawati, Evina},
  title = {Effectiveness of Using a Digital Wearable Plantar Pressure Device to Detect Muscle Fatigue: Within-Subject, Repeated Measures Experimental Design},
  journal = {JMIR Human Factors},
  year = {2025},
  doi = {10.2196/65578}
}
Start for free →