Three Contactless Sleep Technologies Compared With Actigraphy and Polysomnography in a Heterogeneous Group of Older Men and Women in a Model of Mild Sleep Disturbance: Sleep Laboratory Study
JMIR mHealth and uHealth · 25.10.2023 · 11 · e46338
DOI: 10.2196/46338 ↗Kurz gefasst
Fachartikel in der Zeitschrift JMIR mHealth and uHealth, online erschienen am 25.10.2023. Verfasst von Kiran K G Ravindran und 6 weiteren Autor:innen. Beteiligte Einrichtung: Surrey Sleep Research Centre, School of Biosciences, Faculty of Health and Medical Sciences, University of Surrey, Guildford, United Kingdom. Hauptthemen laut MEDLINE: Actigraphy, Sleep. Die Arbeit wurde bisher 39-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Medical Research Council.
Themen: Schlaf
Abstract (Originalsprache)
Background
Contactless sleep technologies (CSTs) hold promise for longitudinal, unobtrusive sleep monitoring in the community and at scale. They may be particularly useful in older populations wherein sleep disturbance, which may be indicative of the deterioration of physical and mental health, is highly prevalent. However, few CSTs have been evaluated in older people.
Objective
This study evaluated the performance of 3 CSTs compared to polysomnography (PSG) and actigraphy in an older population.
Methods
Overall, 35 older men and women (age: mean 70.8, SD 4.9 y; women: n=14, 40%), several of whom had comorbidities, including sleep apnea, participated in the study. Sleep was recorded simultaneously using a bedside radar (Somnofy [Vital Things]: n=17), 2 undermattress devices (Withings sleep analyzer [WSA; Withings Inc]: n=35; Emfit-QS [Emfit; Emfit Ltd]: n=17), PSG (n=35), and actigraphy (Actiwatch Spectrum [Philips Respironics]: n=18) during the first night in a 10-hour time-in-bed protocol conducted in a sleep laboratory. The devices were evaluated through performance metrics for summary measures and epoch-by-epoch classification. PSG served as the gold standard.
Results
The protocol induced mild sleep disturbance with a mean sleep efficiency (SEFF) of 70.9% (SD 10.4%; range 52.27%-92.60%). All 3 CSTs overestimated the total sleep time (TST; bias: >90 min) and SEFF (bias: >13%) and underestimated wake after sleep onset (bias: >50 min). Sleep onset latency was accurately detected by the bedside radar (bias: 16 min). CSTs did not perform as well as actigraphy in estimating the all-night sleep summary measures. In an epoch-by-epoch concordance analysis, the bedside radar performed better in discriminating sleep versus wake (Matthew correlation coefficient [MCC]: mean 0.63, SD 0.12, 95% CI 0.57-0.69) than the undermattress devices (MCC of WSA: mean 0.41, SD 0.15, 95% CI 0.36-0.46; MCC of Emfit: mean 0.35, SD 0.16, 95% CI 0.26-0.43). The accuracy of identifying rapid eye movement and light sleep was poor across all CSTs, whereas deep sleep (ie, slow wave sleep) was predicted with moderate accuracy (MCC: >0.45) by both Somnofy and WSA. The deep sleep duration estimates of Somnofy correlated (r2=0.60; P
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Surrey Sleep Research Centre, School of Biosciences, Faculty of Health and Medical Sciences, University of Surrey, Guildford, United Kingdom.
- UK Dementia Research Institute, Care Research and Technology Centre at Imperial College, London and the University of Surrey, Guildford, United Kingdom.
- Surrey Clinical Research Facility, School of Biosciences, Faculty of Health and Medical Sciences, Guildford, United Kingdom.
- National Institute for Health Research - Royal Surrey Clinical Research Facility, Guildford, United Kingdom.
ORCID-Profile: Kiran K G Ravindran, Ciro Della Monica, Giuseppe Atzori, Damion Lambert, Hana Hassanin, Victoria Revell, Derk-Jan Dijk
Förderung
- Medical Research Council
Zitieren
APA
G Ravindran, K. K., Della Monica, C., Atzori, G., Lambert, D., Hassanin, H., Revell, V., & Dijk, D. J. (2023). Three Contactless Sleep Technologies Compared With Actigraphy and Polysomnography in a Heterogeneous Group of Older Men and Women in a Model of Mild Sleep Disturbance: Sleep Laboratory Study. JMIR mHealth and uHealth, 11, e46338. https://doi.org/10.2196/46338
Vancouver
G Ravindran KK, Della Monica C, Atzori G, Lambert D, Hassanin H, Revell V, et al. Three Contactless Sleep Technologies Compared With Actigraphy and Polysomnography in a Heterogeneous Group of Older Men and Women in a Model of Mild Sleep Disturbance: Sleep Laboratory Study. JMIR mHealth and uHealth. 2023;11:e46338. doi:10.2196/46338
BibTeX
@article{GRavindran2023,
author = {G Ravindran, Kiran K and Della Monica, Ciro and Atzori, Giuseppe and Lambert, Damion and Hassanin, Hana and Revell, Victoria and Dijk, Derk-Jan},
title = {Three Contactless Sleep Technologies Compared With Actigraphy and Polysomnography in a Heterogeneous Group of Older Men and Women in a Model of Mild Sleep Disturbance: Sleep Laboratory Study},
journal = {JMIR mHealth and uHealth},
year = {2023},
volume = {11},
pages = {e46338},
doi = {10.2196/46338}
}