Effects of auxiliary devices for intraoral scanning on the accuracy of scan bodies and implant positioning accuracy for prosthetic design
Journal of Prosthodontic Research · 2026 · 70 (4) · 591-603
DOI: 10.2186/jpr.jpr_d_25_00191 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Prosthodontic Research, 2026. Verfasst von Gyu-Ri Kim und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Dental Science, Graduate School, Kyungpook National University, Daegu, Republic of Korea. Hauptthemen laut MEDLINE: Dental Implants, Dental Prosthesis Design, Computer-Aided Design, Imaging, Three-Dimensional. Die Arbeit nennt 49 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: This study evaluated the effects of auxiliary devices on the scanning accuracy of scan bodies and implant positioning accuracy for dental prosthetic design. METHODS: Four implants were placed in an edentulous maxillary model and fitted with scan bodies. Two types of removable auxiliary devices were designed and 3D-printed for intraoral scanning, categorized into three groups: non-auxiliary device (NAD), geometric-shaped device (GAD), and tooth-shaped device (TAD) groups. Scanning was performed using three intraoral scanners (i700, i500, and Primescan), with ten scans acquired per group. The scanning accuracy of scan bodies was assessed using 3D inspection software (Zeiss Inspect; Zeiss), whereas implant positioning accuracy after scan body library alignment was analyzed using Geomagic Control X (3D Systems) and Zeiss Inspect. Statistical analyses were performed using one-way and two-way ANOVA followed by Tukey HSD tests to compare the performance among the three groups (α = 0.05). RESULTS: The TAD group demonstrated significantly lower linear discrepancy in the anterior and posterior scan body distance for the i700 and i500 (P < 0.001). However, in terms of implant position accuracy after scan body library alignment, the NAD group with the i700 system showed significantly lower root mean square (RMS) values (P = 0.001) and also demonstrated significantly lower deviation in the apical region (P = 0.005). CONCLUSIONS: The use of auxiliary devices for intraoral scanning improved the accuracy of the scan body; however, these devices had little or even a detrimental effect on implant positioning accuracy after scan body library alignment in the CAD process.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Dental Science, Graduate School, Kyungpook National University, Daegu, Republic of Korea.
- Advanced Dental Device Development Institute (A3DI), Kyungpook National University, Daegu, Republic of Korea.
- Department of Prosthodontics, School of Dentistry, Kyungpook National University, Daegu, Republic of Korea.
Zitieren
APA
Kim, G. R., Son, K., Kim, S. Y., & Lee, K. B. (2026). Effects of auxiliary devices for intraoral scanning on the accuracy of scan bodies and implant positioning accuracy for prosthetic design. Journal of Prosthodontic Research, 70(4), 591-603. https://doi.org/10.2186/jpr.jpr_d_25_00191
Vancouver
Kim GR, Son K, Kim SY, Lee KB. Effects of auxiliary devices for intraoral scanning on the accuracy of scan bodies and implant positioning accuracy for prosthetic design. Journal of Prosthodontic Research. 2026;70(4):591-603. doi:10.2186/jpr.jpr_d_25_00191
BibTeX
@article{Kim2026,
author = {Kim, Gyu-Ri and Son, KeunBaDa and Kim, So-Yeun and Lee, Kyu-Bok},
title = {Effects of auxiliary devices for intraoral scanning on the accuracy of scan bodies and implant positioning accuracy for prosthetic design},
journal = {Journal of Prosthodontic Research},
year = {2026},
volume = {70},
number = {4},
pages = {591--603},
doi = {10.2186/jpr.jpr_d_25_00191}
}
EUVIMED