Cuff inflation during ambulatory blood pressure monitoring and heart rate

Mia Skov-Madsen, My Svensson, Jeppe Hagstrup Christensen

Integrated Blood Pressure Control · 2008-11 · Volume 1 · 15-19

DOI: 10.2147/ibpc.s4072 ↗

At a glance

Journal article in Integrated Blood Pressure Control, 2008-11. Written by Mia Skov-Madsen and 2 co-authors. Institution: Department of Nephrology, Aarhus University Hospital, Aalborg, Denmark. Cited 2 times so far (Crossref). The full text is freely available (open access, CC BY-NC 3.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

INTRODUCTION: Twenty four-hour ambulatory blood pressure monitoring is a clinically validated procedure in evaluation of blood pressure (BP). We hypothesised that the discomfort during cuff inflation would increase the heart rate (HR) measured with 24-h ambulatory BP monitoring compared to a following HR measurement with a 24-h Holter monitor. METHODS: The study population (n = 56) were recruited from the outpatient's clinic at the Department of Nephrology, Aalborg Hospital, Aarhus University Hospital at Aalborg, Denmark. All the patients had chronic kidney disease (CKD). We compared HR measured with a 24-h Holter monitor with a following HR measured by a 24-h ambulatory BP monitoring. RESULTS: We found a highly significant correlation between the HR measured with the Holter monitor and HR measured with 24-h ambulatory blood pressure monitoring (r = 0.77, p < 0.001). Using the Bland-Altman plot, the mean difference in HR was only 0.5 beat/min during 24 hours with acceptable limits of agreement for both high and low HR levels. Dividing the patients into groups according to betablocker treatment, body mass index, age, sex, angiotensin-converting enzyme inhibitor treatment, statins treatment, diuretic treatment, or calcium channel blocker treatment revealed similar results as described above. CONCLUSION: The results indicate that the discomfort induced by cuff inflation during 24-h ambulatory BP monitoring does not increase HR. Thus, 24-h ambulatory BP monitoring may be a reliable measurement of the BP among people with CKD.

Subjects

Author keywords

Holter monitoringambulatory blood pressure monitoringchronic kidney diseaseheart ratehypertension

Authors and institutions

  • Department of Nephrology, Aarhus University Hospital, Aalborg, Denmark.

Cite

APA

Skov-Madsen, M., Svensson, M., & Christensen, J. H. (2008). Cuff inflation during ambulatory blood pressure monitoring and heart rate. Integrated Blood Pressure Control, Volume 1, 15-19. https://doi.org/10.2147/ibpc.s4072

Vancouver

Skov-Madsen M, Svensson M, Christensen JH. Cuff inflation during ambulatory blood pressure monitoring and heart rate. Integrated Blood Pressure Control. 2008;Volume 1:15-19. doi:10.2147/ibpc.s4072

BibTeX
@article{SkovMadsen2008,
  author = {Skov-Madsen, Mia and Svensson, My and Christensen, Jeppe Hagstrup},
  title = {Cuff inflation during ambulatory blood pressure monitoring and heart rate},
  journal = {Integrated Blood Pressure Control},
  year = {2008},
  volume = {Volume 1},
  pages = {15--19},
  doi = {10.2147/ibpc.s4072}
}
Start for free →