Fallbericht in der Zeitschrift JBJS Case Connector, 2026-07. Verfasst von Jacqueline R Ray und 4 weiteren Autor:innen. Beteiligte Einrichtung: Anderson Orthopedic Research Institute, Alexandria, Virginia. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Hip Prosthesis, Prosthesis Failure. Die Arbeit nennt 10 Literaturangaben.
Abstract (Originalsprache)
Case: A 79-year-old man presented with progressive right thigh pain over 6 months and reliance on a cane. He had undergone primary total hip arthroplasty (THA) with a metal-on-metal bearing in 2007. In 2025, radiographs revealed a fracture of his 18-mm diameter extensively porous-coated, cobalt-chrome alloy femoral stem (Prodigy, DePuy Orthopaedics), prompting revision THA. To the best of our knowledge, this case represents the largest diameter fracture of this type of component reported to date. Conclusion: While fractures of smaller diameter stems are well-documented, clinicians should be aware that fatigue fractures can also occur in very large diameter components with a cylindrical distal geometry.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Anderson Orthopedic Research Institute, Alexandria, Virginia.
ORCID-Profile: Jacqueline R Ray, Christopher M Jaicks, Henry Ho, Robert H Hopper
Zitieren
APA
Ray, J. R., Jaicks, C. M., Ho, H., Hopper, R. H., & McAsey, C. J. (2026). Fracture of an 18-mm Diameter Extensively Porous-Coated Femoral Stem 17.9 Years After Primary Total Hip Arthroplasty. JBJS Case Connector, 16(3). https://doi.org/10.2106/jbjs.cc.26.00127
Vancouver
Ray JR, Jaicks CM, Ho H, Hopper RH, McAsey CJ. Fracture of an 18-mm Diameter Extensively Porous-Coated Femoral Stem 17.9 Years After Primary Total Hip Arthroplasty. JBJS Case Connector. 2026;16(3). doi:10.2106/jbjs.cc.26.00127
BibTeX
@article{Ray2026,
author = {Ray, Jacqueline R and Jaicks, Christopher M and Ho, Henry and Hopper, Robert H and McAsey, Craig J},
title = {Fracture of an 18-mm Diameter Extensively Porous-Coated Femoral Stem 17.9 Years After Primary Total Hip Arthroplasty},
journal = {JBJS Case Connector},
year = {2026},
volume = {16},
number = {3},
doi = {10.2106/jbjs.cc.26.00127}
}
EUVIMED