Fallbericht in der Zeitschrift JBJS Case Connector, 2026-07. Verfasst von Jacqueline R Ray und 4 weiteren Autor:innen. Beteiligte Einrichtung: Anderson Orthopedic Research Institute, Alexandria, Virginia. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Hip Prosthesis, Prosthesis Failure. Die Arbeit nennt 10 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Case: A 79-year-old man presented with progressive right thigh pain over 6 months and reliance on a cane. He had undergone primary total hip arthroplasty (THA) with a metal-on-metal bearing in 2007. In 2025, radiographs revealed a fracture of his 18-mm diameter extensively porous-coated, cobalt-chrome alloy femoral stem (Prodigy, DePuy Orthopaedics), prompting revision THA. To the best of our knowledge, this case represents the largest diameter fracture of this type of component reported to date. Conclusion: While fractures of smaller diameter stems are well-documented, clinicians should be aware that fatigue fractures can also occur in very large diameter components with a cylindrical distal geometry.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, HipHip ProsthesisProsthesis FailureHumansMaleAgedReoperationProsthesis Design

Stichwörter der Autor:innen

extensively porous-coated cylindrical stem fatigue fracturelarge diameter femoral stemmetal-on-metal bearingtotal hip arthroplasty

Autor:innen und Institutionen

  • Anderson Orthopedic Research Institute, Alexandria, Virginia.

ORCID-Profile: Jacqueline R Ray, Christopher M Jaicks, Henry Ho, Robert H Hopper

Zitieren

APA

Ray, J. R., Jaicks, C. M., Ho, H., Hopper, R. H., & McAsey, C. J. (2026). Fracture of an 18-mm Diameter Extensively Porous-Coated Femoral Stem 17.9 Years After Primary Total Hip Arthroplasty. JBJS Case Connector, 16(3). https://doi.org/10.2106/jbjs.cc.26.00127

Vancouver

Ray JR, Jaicks CM, Ho H, Hopper RH, McAsey CJ. Fracture of an 18-mm Diameter Extensively Porous-Coated Femoral Stem 17.9 Years After Primary Total Hip Arthroplasty. JBJS Case Connector. 2026;16(3). doi:10.2106/jbjs.cc.26.00127

BibTeX
@article{Ray2026,
  author = {Ray, Jacqueline R and Jaicks, Christopher M and Ho, Henry and Hopper, Robert H and McAsey, Craig J},
  title = {Fracture of an 18-mm Diameter Extensively Porous-Coated Femoral Stem 17.9 Years After Primary Total Hip Arthroplasty},
  journal = {JBJS Case Connector},
  year = {2026},
  volume = {16},
  number = {3},
  doi = {10.2106/jbjs.cc.26.00127}
}
Kostenlos starten →