Autoclaving Effectively Sterilizes Contaminated Retained Cement in Orthopaedic Surgical Trays

Andrew L Thomson iD, Christina Chao iD, Mohammad Hammad iD, Mario Mendia, Mathias P G Bostrom iD, Alberto V Carli

Journal of Bone and Joint Surgery · 2026

DOI: 10.2106/jbjs.25.00655 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Bone and Joint Surgery, 2026. Verfasst von Andrew L Thomson und 5 weiteren Autor:innen. Beteiligte Einrichtung: Arthroplasty Research Laboratory, Research Institute, Hospital for Special Surgery, New York, NY. Hauptthemen laut MEDLINE: Polymethyl Methacrylate, Sterilization, Equipment Contamination, Bone Cements. Die Arbeit nennt 24 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Retained polymethylmethacrylate (PMMA) in surgical instrument trays and on instruments is becoming more common and is a disquieting situation for the arthroplasty operating room team. Although retained debris may be sterile after autoclaving, previous studies have not specifically investigated bacterial viability on retained PMMA. This study determined whether heavily contaminated PMMA surfaces and instruments covered in PMMA could be effectively sterilized by clinically utilized autoclaving protocols.

Methods

Mature methicillin-susceptible Staphylococcus aureus (MSSA) and Escherichia coli biofilms were grown on PMMA coupons with a smooth surface (n = 5 per condition) or with deep crevices (n = 5 per condition) and on threaded screws, which were inserted into polymerizing PMMA (n = 5 per condition). PMMA coupons and PMMA-covered screws were subjected to 3 autoclave sterilization protocols. Colony-forming unit (CFU) counts and adenosine triphosphate (ATP) levels were measured to detect viable bacteria. Scanning electron microscope (SEM) images were obtained to determine whether biofilm was present.

Results

Non-autoclaved PMMA surfaces were heavily contaminated, with values of 4.1 ± 2.3 × 10 6 CFUs/mL for S . aureus and 2.4 ± 1.2 × 10 6 CFUs/mL for E . coli. ATP levels of 3,230 ± 1,297 relative light units (RLU) for S . aureus and 1,006 ± 602 RLU for E . coli were observed. Crevices and screws were also heavily contaminated. Following autoclaving, CFUs/mL counts of

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Polymethyl MethacrylateSterilizationEquipment ContaminationBone CementsBiofilmsEscherichia coliStaphylococcus aureusHumansColony Count, MicrobialMicroscopy, Electron, Scanning

Substanzen

Polymethyl MethacrylateBone Cements

Autor:innen und Institutionen

  • Arthroplasty Research Laboratory, Research Institute, Hospital for Special Surgery, New York, NY.
  • Department of Orthopedic Surgery, Hospital for Special Surgery, New York, NY.

ORCID-Profile: Andrew L Thomson, Christina Chao, Mohammad Hammad, Mathias P G Bostrom

Zitieren

APA

Thomson, A. L., Chao, C., Hammad, M., Mendia, M., Bostrom, M. P. G., & Carli, A. V. (2026). Autoclaving Effectively Sterilizes Contaminated Retained Cement in Orthopaedic Surgical Trays. Journal of Bone and Joint Surgery. https://doi.org/10.2106/jbjs.25.00655

Vancouver

Thomson AL, Chao C, Hammad M, Mendia M, Bostrom MPG, Carli AV. Autoclaving Effectively Sterilizes Contaminated Retained Cement in Orthopaedic Surgical Trays. Journal of Bone and Joint Surgery. 2026. doi:10.2106/jbjs.25.00655

BibTeX
@article{Thomson2026,
  author = {Thomson, Andrew L and Chao, Christina and Hammad, Mohammad and Mendia, Mario and Bostrom, Mathias P G and Carli, Alberto V},
  title = {Autoclaving Effectively Sterilizes Contaminated Retained Cement in Orthopaedic Surgical Trays},
  journal = {Journal of Bone and Joint Surgery},
  year = {2026},
  doi = {10.2106/jbjs.25.00655}
}
Kostenlos starten →