Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Bone and Joint Surgery, online erschienen am 21.02.2024. Verfasst von Yong-Jun Lee und 5 weiteren Autor:innen. Beteiligte Einrichtung: Arthroscopy and Joint Research Institute, Department of Orthopaedic Surgery, Severance Hospital, Yonsei University College of Medicine, Seoul, Republ…. Hauptthemen laut MEDLINE: Bone Plates, Bone Screws, Periprosthetic Fractures, Fracture Fixation, Internal. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 25 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

In the setting of periprosthetic humeral fractures, the humeral stem of the implant represents a substantial challenge to the optimal method of proximal fixation. This study aimed to compare the initial biomechanical stability provided by cerclage cables with a locking plate insert versus bicortical locking screws (i.e., the gold standard for fixation) in fresh cadaveric humeri.

Methods

After calculating the sample size, we utilized 10 sets of cadaveric specimens and created a 5-mm osteotomy gap 120 mm distal to the tip of the greater tuberosity, simulating a Wright and Cofield type-B periprosthetic humeral fracture on each specimen. Using 3 locking screws for distal fragment fixation, identical in all specimens, the specimens were assigned to Group A (3 cerclage cables with a plate insert) or Group B (3 locking bicortical screws) for proximal fragment fixation. Biomechanical tests included stiffness in varus and valgus bending, torsion, and axial compression, and a single load to failure.

Results

No significant differences were observed in the biomechanical metrics between the 2 groups.

Conclusions

Our study revealed that fixation with use of cerclage cables with a plate insert demonstrated biomechanical stability comparable with that of bicortical locking screw fixation when addressing the proximal fragment in Wright and Cofield type-B periprosthetic humeral fractures. Clinical Relevance: For proximal fragment fixation of periprosthetic humeral fractures, cerclage cables with a plate insert can be utilized as an effective fixation method that offers initial fixation strength that is comparable to the use of 3 locking bicortical screws.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Bone PlatesBone ScrewsPeriprosthetic FracturesFracture Fixation, InternalCadaverHumansBiomechanical PhenomenaAgedFemaleMaleHumeral FracturesAged, 80 and over

Autor:innen und Institutionen

  • Arthroscopy and Joint Research Institute, Department of Orthopaedic Surgery, Severance Hospital, Yonsei University College of Medicine, Seoul, Republic of Korea.

ORCID-Profile: Yong-Jun Lee, Joon-Ryul Lim, Jee-Hoon Choi, Tae-Hwan Yoon, Yun-Rak Choi, Yong-Min Chun

Zitieren

APA

Lee, Y. J., Lim, J. R., Choi, J. H., Yoon, T. H., Choi, Y. R., & Chun, Y. M. (2024). Comparing the Biomechanical Stability of Cerclage Cable with Plate Insert Versus Locking Screw in Periprosthetic Humeral Fracture. Journal of Bone and Joint Surgery, 106(9), 817-822. https://doi.org/10.2106/jbjs.23.00830

Vancouver

Lee YJ, Lim JR, Choi JH, Yoon TH, Choi YR, Chun YM. Comparing the Biomechanical Stability of Cerclage Cable with Plate Insert Versus Locking Screw in Periprosthetic Humeral Fracture. Journal of Bone and Joint Surgery. 2024;106(9):817-822. doi:10.2106/jbjs.23.00830

BibTeX
@article{Lee2024,
  author = {Lee, Yong-Jun and Lim, Joon-Ryul and Choi, Jee-Hoon and Yoon, Tae-Hwan and Choi, Yun-Rak and Chun, Yong-Min},
  title = {Comparing the Biomechanical Stability of Cerclage Cable with Plate Insert Versus Locking Screw in Periprosthetic Humeral Fracture},
  journal = {Journal of Bone and Joint Surgery},
  year = {2024},
  volume = {106},
  number = {9},
  pages = {817--822},
  doi = {10.2106/jbjs.23.00830}
}