Application of metal artifact reduction algorithm in reducing metal artifacts in post-surgery pediatric low radiation dose spine computed tomography (CT) images
Quantitative Imaging in Medicine and Surgery · 2024-07 · 14 (7) · 4648-4658
DOI: 10.21037/qims-23-1659 ↗Kurz gefasst
Fachartikel in der Zeitschrift Quantitative Imaging in Medicine and Surgery, 2024-07. Verfasst von Jihang Sun und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Beijing Children's Hospital, Capital Medical University, National Center for Children's Health, Beijing, China. Die Arbeit wurde bisher 6-mal zitiert (Crossref). Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
The commonly used methods for removing metal-induced beam hardening artifacts often rely on the use of high energy photons with either high tube voltage or high energy virtual monoenergetic images in dual-energy computed tomography (CT), the radiation dose was usually relatively high in order to generate adequate signals. This retrospective study is designed to evaluate the application of a metal artifact reduction (MAR) algorithm in reducing pedicle screw metal-caused beam hardening artifacts in post-surgery pediatric low radiation dose spine CT images.
Methods
Seventy-seven children (3-15 years) who had undergone a low dose CT with 140 or 100 kV were enrolled. The radiation dose was 1.40 mGy for the 3-8 years old and 2.61 mGy for 9-15 years old children. There were 116 pedicle screws evaluated. The raw data were reconstructed with adaptive statistical iterative reconstruction-V (ASIR-V) at 50% strength, ASIR-V with MAR (AV-MAR), deep learning image reconstruction (DLIR) at high strength and DLIR with MAR (DL-MAR). The image quality concerning pedicle screws was evaluated objectively in terms of the length of beam-hardening artifact (LHA) and artifact index (AI), and subjectively using a 4-point scale (4 points: best, 3 points: acceptable).
Results
Both AV-MAR and DL-MAR significantly reduced metal-induced beam hardening artifacts with smaller LHA (15.76±10.12 mm, a reduction of 57.24% and 15.66±10.49 mm, a reduction of 57.40%, respectively), and AI value (62.50±33.51, a reduction of 64.65% and 61.03±32.61, a reduction of 65.01%, respectively) compared to ASIR-V and DLIR (all P
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Beijing Children's Hospital, Capital Medical University, National Center for Children's Health, Beijing, China.
- Medical Imaging Department, Children's Hospital of Xinjiang Uygur Autonomous Region, Xinjiang Hospital of Beijing Children's Hospital, Urumqi, China.
- Department of Radiology, Fujian Maternity and Child Health Hospital College of Clinical Medicine for Obstetrics & Gynecology and Pediatrics, Fujian Medical University, Fujian, China.
Zitieren
APA
Sun, J., Li, H., Yu, T., Huo, A., Hua, S., Zhou, Z., & Peng, Y. (2024). Application of metal artifact reduction algorithm in reducing metal artifacts in post-surgery pediatric low radiation dose spine computed tomography (CT) images. Quantitative Imaging in Medicine and Surgery, 14(7), 4648-4658. https://doi.org/10.21037/qims-23-1659
Vancouver
Sun J, Li H, Yu T, Huo A, Hua S, Zhou Z, et al. Application of metal artifact reduction algorithm in reducing metal artifacts in post-surgery pediatric low radiation dose spine computed tomography (CT) images. Quantitative Imaging in Medicine and Surgery. 2024;14(7):4648-4658. doi:10.21037/qims-23-1659
BibTeX
@article{Sun2024,
author = {Sun, Jihang and Li, Haoyan and Yu, Tong and Huo, Aihua and Hua, Shan and Zhou, Zuofu and Peng, Yun},
title = {Application of metal artifact reduction algorithm in reducing metal artifacts in post-surgery pediatric low radiation dose spine computed tomography (CT) images},
journal = {Quantitative Imaging in Medicine and Surgery},
year = {2024},
volume = {14},
number = {7},
pages = {4648--4658},
doi = {10.21037/qims-23-1659}
}
EUVIMED