Hemoglobin Shelby interfered with the leukocyte differential channel of the Mindray BC-6800Plus hematology analyzer: a case report
Journal of Laboratory and Precision Medicine · 2022-04 · 7 · 14-14
DOI: 10.21037/jlpm-21-56 ↗Kurz gefasst
Fallbericht in der Zeitschrift Journal of Laboratory and Precision Medicine, 2022-04. Verfasst von Eloísa Urrechaga und 4 weiteren Autor:innen. Die Arbeit wurde bisher 2-mal zitiert (Crossref).
Abstract (Originalsprache)
Background
Hemoglobin (Hb) variants are the world’s most frequent monogenic disorders; up to 150 are unstable. Mild hemolytic anemia can be the only clinical evidence that a patient has unstable Hb variants, but the minimal changes in complete blood count (CBC) are difficult to observe and often mean that the condition is undetected in most cases. Case: Description: We present a report of a Hb Shelby carrier. An asymptomatic 47-year-old woman born in Bolivia was referred to our medical center for a routine health control. CBC was analyzed with a Mindray BC6800 plus counter and showed values within reference ranges. Only reticulocytosis was detected. The white blood cell differential channel showed an abnormally low fluorescent signal in the leukocyte (WBC) differential scattergram. This pattern has been observed in other counters based on similar fluorescence detection principles, linked to the presence of Hb variants, so reflect tests were performed to confirm this clinical condition.
Conclusions
Incidental findings in CBC are not uncommon, an abnormal WBC scatterplot must be investigated and can be used to detect carriers in order to reach the diagnosis efficiently.
Zitieren
APA
Urrechaga, E., Merino, M., Fernández, M., Aguirrezabal, J., & Orbe, R. D. (2022). Hemoglobin Shelby interfered with the leukocyte differential channel of the Mindray BC-6800Plus hematology analyzer: a case report. Journal of Laboratory and Precision Medicine, 7, 14-14. https://doi.org/10.21037/jlpm-21-56
Vancouver
Urrechaga E, Merino M, Fernández M, Aguirrezabal J, Orbe RD. Hemoglobin Shelby interfered with the leukocyte differential channel of the Mindray BC-6800Plus hematology analyzer: a case report. Journal of Laboratory and Precision Medicine. 2022;7:14-14. doi:10.21037/jlpm-21-56
BibTeX
@article{Urrechaga2022,
author = {Urrechaga, Eloísa and Merino, Mónica and Fernández, Mónica and Aguirrezabal, Jaione and Orbe, Rafael del},
title = {Hemoglobin Shelby interfered with the leukocyte differential channel of the Mindray BC-6800Plus hematology analyzer: a case report},
journal = {Journal of Laboratory and Precision Medicine},
year = {2022},
volume = {7},
pages = {14--14},
doi = {10.21037/jlpm-21-56}
}