EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Mechanistic investigation and data-driven correction of lipemic interference in hematological parameters on the Mindray BC-6800 hematology analyzer

Shoumin Li, Haiping Yang, Qian Zhang, Fu Zhao, Yong Mu, Yingjiang Chen

Practical Laboratory Medicine · 2026-04 · 49 · e00521

DOI: 10.1016/j.plabm.2026.e00521 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Practical Laboratory Medicine, 2026-04. Verfasst von Shoumin Li und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Laboratory, Liupanshui Municipal People's Hospital, Liupanshui, 553001, China. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

This study aimed to systematically investigate the interference of lipemia on the Mindray BC-6800 hematology analyzer, elucidate the underlying mechanisms via optical signal analysis, and develop data-driven strategies for its recognition and correction.

Methods

In vitro models of moderate and severe lipemia were established using 30 healthy specimens. Complete blood count (CBC) parameters, leukocyte differentials, and optical signals were measured and compared against controls. A random forest model was constructed for interference identification, and a multiple linear regression model was developed from 600 normal samples to predict hemoglobin (HGB).

Results

Lipemia significantly increased Hb, MCH, and MCHC (P 36 pg & MCHC >410 g/L″) attained 81.1% accuracy. The HGB prediction model (HGB = 6.67 + 3.14 × HCT) performed robustly across lipemia levels (R2 > 0.96).

Conclusions

Lipemia interferes with multiple parameters on the Mindray BC-6800 through alterations in cellular optical signals. The developed recognition and correction models demonstrate high clinical applicability for improving result accuracy.

Schlagwörter

Stichwörter der Autor:innen

Analytical interferenceComplete blood countLipemiaMachine learningMindray BC-6800Prediction model

Autor:innen und Institutionen

  • Department of Clinical Laboratory, Liupanshui Municipal People's Hospital, Liupanshui, 553001, China.
  • School of Laboratory Medicine, Guizhou Medical University, Guiyang, 550004, China.
  • Department of Orthopedics and Sports Medicine, Liupanshui Municipal People's Hospital, Liupanshui, 553001, China.

Zitieren

APA

Li, S., Yang, H., Zhang, Q., Zhao, F., Mu, Y., & Chen, Y. (2026). Mechanistic investigation and data-driven correction of lipemic interference in hematological parameters on the Mindray BC-6800 hematology analyzer. Practical Laboratory Medicine, 49, e00521. https://doi.org/10.1016/j.plabm.2026.e00521

Vancouver

Li S, Yang H, Zhang Q, Zhao F, Mu Y, Chen Y. Mechanistic investigation and data-driven correction of lipemic interference in hematological parameters on the Mindray BC-6800 hematology analyzer. Practical Laboratory Medicine. 2026;49:e00521. doi:10.1016/j.plabm.2026.e00521

BibTeX
@article{Li2026,
  author = {Li, Shoumin and Yang, Haiping and Zhang, Qian and Zhao, Fu and Mu, Yong and Chen, Yingjiang},
  title = {Mechanistic investigation and data-driven correction of lipemic interference in hematological parameters on the Mindray BC-6800 hematology analyzer},
  journal = {Practical Laboratory Medicine},
  year = {2026},
  volume = {49},
  pages = {e00521},
  doi = {10.1016/j.plabm.2026.e00521}
}
Kostenlos starten →