Decreased Pulse Oximetry Readings in Asymptomatic Patient with Hemoglobin Grifton
Anesthesia eJournal · 2021
DOI: 10.18776/6sd50546 ↗Kurz gefasst
Fachartikel in der Zeitschrift Anesthesia eJournal, 2021. Verfasst von Anne Marie Serrano.
Abstract (Originalsprache)
Hemoglobin Grifton presents as a decreased oxygen saturation as measured by pulse oximetry (SpO2) reading despite a normal arterial oxygen saturation (SaO2) in an asymptomatic patient. Hemoglobin Grifton is caused by a mutation on an alpha chain of hemoglobin at codon 87, which affects the absorption wavelength of its oxyhemoglobin. A standard pulse oximeter is designed to measure the absorption of light at specific wavelengths and perceives the hemoglobin Grifton oxyhemoglobin as a deoxyhemoglobin. In a patient with a known variant hemoglobin, ordering additional diagnostic or therapeutic testing can increase stress for a patient requiring routine medical interventions or emergencies. Careful evaluation of each individual case is warranted prior to proceeding with an anesthetic.
Zitieren
APA
Serrano, A. M. (2021). Decreased Pulse Oximetry Readings in Asymptomatic Patient with Hemoglobin Grifton. Anesthesia eJournal. https://doi.org/10.18776/6sd50546
Vancouver
Serrano AM. Decreased Pulse Oximetry Readings in Asymptomatic Patient with Hemoglobin Grifton. Anesthesia eJournal. 2021. doi:10.18776/6sd50546