EUVIMEDEUVIMED entdecken
Open Access

Novel SOI-Biosensor Topology for the Detection of an Acute Myocardial Infarction Marker — Troponin I

A A Cheremiskina, V V Krasitskaya, V M Generalov, L A Frank, A V Glukhov, M V Kruchinina, G A Kudrov, D E Serdyuk, V K Grabezhova

Sovremennye tehnologii v medicine · 2024

DOI: 10.17691/stm2024.16.1.04 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sovremennye tehnologii v medicine, 2024. Verfasst von A A Cheremiskina und 8 weiteren Autor:innen. Beteiligte Einrichtung: Junior Researcher; Federal Budgetary Research Institution, State Research Center of Virology and Biotechnology "Vector", Federal Service for Surveill…. Hauptthemen laut MEDLINE: Troponin I, Biosensing Techniques, Myocardial Infarction, Aptamers, Nucleotide. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A biosensor based on field-effect transistors on silicon-on-insulator structures (SOI-biosensor) is a high-potential device for detection of biological molecules, for instance, such as troponin I; the biosensor allows conducting label-free real-time analysis. The aim of the study is the development of SOI-biosensor design for detection of acute myocardial infarction marker - troponin I. A notable feature of this design was the integration of two grounding electrodes directly onto the biosensor surface, which effectively nullified the static potential of the liquid sample and minimized physical breakdowns of biosensor elements. Materials and Methods: The highly specific anti-troponin I DNA aptamer was used as a receptor for specific detection of protein marker. Aptamer immobilization on the biosensor surface was carried out by physical adsorption. The analyzed range of target troponin I molecules concentration in the sample varied within 10-11 to 10-9 mol/L, mirroring clinical levels observed in myocardial infarction cases. During the experiment, a constant voltage of Vds=0.15 V was maintained. Results: The developed SOI-biosensor successfully detected target troponin I molecules at a concentration of 10-11 mol/L. The detection process exhibited an effective time of approximately 200-300 s per sample. Moreover, analysis of the detection process revealed a noticeable decrease in current within the source-drain circuit, indicative of the negatively charged complex formed by troponin I and anti-troponin I DNA-aptamer at the "liquid sample-nanowire" phase interface.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Troponin IBiosensing TechniquesMyocardial InfarctionAptamers, NucleotideBiomarkersHumansTransistors, Electronic

Stichwörter der Autor:innen

aptamerbiosensorfield-effect transistormyocardial infarctionnanowirephysical adsorptionsilicon-on-insulatortroponin I

Substanzen

Troponin IAptamers, NucleotideBiomarkers

Autor:innen und Institutionen

  • Junior Researcher; Federal Budgetary Research Institution, State Research Center of Virology and Biotechnology "Vector", Federal Service for Surveillance on Consumer Rights Protection and Human Wellbeing, Koltsovo, Novo…
  • Senior Researcher; Institute of Biophysics of the Siberian Branch of the Russian Academy of Sciences, Federal Research Center "Krasnoyarsk Science Center of the Siberian Branch of the Russian Academy of Sciences", 50/50…
  • Leading Researcher; Federal Budgetary Research Institution, State Research Center of Virology and Biotechnology "Vector", Federal Service for Surveillance on Consumer Rights Protection and Human Wellbeing, Koltsovo, Nov…
  • Chief Researcher; Institute of Biophysics of the Siberian Branch of the Russian Academy of Sciences, Federal Research Center "Krasnoyarsk Science Center of the Siberian Branch of the Russian Academy of Sciences", 50/50…
  • Deputy General Director for Research; Joint Stock Company "Novosibirsk Factory of Semiconductor Devices VOSTOK", 60 Dachnaya St., Novosibirsk, 630082, Russia.
  • Associate Professor, Leading Researcher; Research Institute of Internal and Preventive Medicine - Branch of the Institute of Cytology and Genetics, Siberian Branch of Russian Academy of Sciences, 175/1 B. Bogatkov St.,…

Zitieren

APA

Cheremiskina, A. A., Krasitskaya, V. V., Generalov, V. M., Frank, L. A., Glukhov, A. V., Kruchinina, M. V., Kudrov, G. A., Serdyuk, D. E., & Grabezhova, V. K. (2024). Novel SOI-Biosensor Topology for the Detection of an Acute Myocardial Infarction Marker — Troponin I. Sovremennye tehnologii v medicine. https://doi.org/10.17691/stm2024.16.1.04

Vancouver

Cheremiskina AA, Krasitskaya VV, Generalov VM, Frank LA, Glukhov AV, Kruchinina MV, et al. Novel SOI-Biosensor Topology for the Detection of an Acute Myocardial Infarction Marker — Troponin I. Sovremennye tehnologii v medicine. 2024. doi:10.17691/stm2024.16.1.04

BibTeX
@article{Cheremiskina2024,
  author = {Cheremiskina, A A and Krasitskaya, V V and Generalov, V M and Frank, L A and Glukhov, A V and Kruchinina, M V and Kudrov, G A and Serdyuk, D E and Grabezhova, V K},
  title = {Novel SOI-Biosensor Topology for the Detection of an Acute Myocardial Infarction Marker — Troponin I},
  journal = {Sovremennye tehnologii v medicine},
  year = {2024},
  doi = {10.17691/stm2024.16.1.04}
}
Kostenlos starten →