Measurement uncertainty in the calibration of a thermal resistor for measuring heat fluxes
tm - Technisches Messen · 24.06.2026 · 93 (5-6) · 192-201
DOI: 10.1515/teme-2026-0019 ↗Kurz gefasst
Fachartikel in der Zeitschrift tm - Technisches Messen, online erschienen am 07.05.2026. Verfasst von Joseph Beerel und 2 weiteren Autor:innen. Die Arbeit nennt 8 Literaturangaben.
Abstract (Originalsprache)
A calibration facility for the calibration of thin plate heat flux sensors following the principle of a comparison calibration has been developed at TU Ilmenau. The reference heat flux sensors employed in this facility require traceable calibration over a temperature range from −30 °C to 70 °C. For this purpose, the thermal resistance of the reference sensors is determined using a modified guarded hot plate configuration. This paper presents the experimental setup used for the calibration of the reference heat flux sensors. Furthermore, the model equation governing the calibration process is derived, and the associated measurement uncertainty is evaluated. Finally, a temperature-dependent characteristic function describing the thermal resistance of the reference sensors is determined. An additional uncertainty analysis yields a relative measurement uncertainty of 2.2 % for the characteristic function, corresponding to a coverage factor of k = 2.
Zitieren
APA
Beerel, J., Vonken, T., & Fröhlich, T. (2026). Measurement uncertainty in the calibration of a thermal resistor for measuring heat fluxes. tm - Technisches Messen, 93(5-6), 192-201. https://doi.org/10.1515/teme-2026-0019
Vancouver
Beerel J, Vonken T, Fröhlich T. Measurement uncertainty in the calibration of a thermal resistor for measuring heat fluxes. tm - Technisches Messen. 2026;93(5-6):192-201. doi:10.1515/teme-2026-0019
BibTeX
@article{Beerel2026,
author = {Beerel, Joseph and Vonken, Tristan and Fröhlich, Thomas},
title = {Measurement uncertainty in the calibration of a thermal resistor for measuring heat fluxes},
journal = {tm - Technisches Messen},
year = {2026},
volume = {93},
number = {5-6},
pages = {192--201},
doi = {10.1515/teme-2026-0019}
}
EUVIMED