Using micro spheres as reference artifacts for the in-situ characterization of tactile 3D micro probes along the probing sphere’s equator
tm - Technisches Messen · 27.05.2024 · 91 (5) · 244-254
DOI: 10.1515/teme-2023-0164 ↗Kurz gefasst
Fachartikel in der Zeitschrift tm - Technisches Messen, online erschienen am 11.03.2024. Verfasst von Erik Oertel und 1 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Gefördert von: Deutsche Forschungsgemeinschaft.
Abstract (Originalsprache)
Nano and micro coordinate measuring machines (CMMs) have been developed for the characterization of small dimensional features. They require a procedure which enables a traceable and precise characterization of probing spheres. In this contribution we explore the use of well characterized micro spheres as reference artifacts for the in-situ characterization of probing spheres along the probing sphere’s equator. The spheres are characterized using a strategy which is based on a set of tactile surface scans in conjunction with a stitching-algorithm. These micro spheres serve as a reference for the in-situ characterization of a tactile 3D micro probe on a nano measuring machine (NMM-1). Our investigations are based on a sample of eight spheres sourced from two different suppliers. Although the sample is small, we could already observe characteristics which seem to be typical for spheres of a certain type (i.e. nominal radius and material). The experiments indicate that micro spheres are a suitable reference artifact for tactile 3D micro probes. We were able to reproduce the measured mean radius of the probing sphere with a standard deviation of 31 nm using reference spheres whose nominal radius covers a range of 89 µm (61 µm up to 150 µm).
Förderung
- Deutsche Forschungsgemeinschaft
Zitieren
APA
Oertel, E., & Manske, E. (2024). Using micro spheres as reference artifacts for the in-situ characterization of tactile 3D micro probes along the probing sphere’s equator. tm - Technisches Messen, 91(5), 244-254. https://doi.org/10.1515/teme-2023-0164
Vancouver
Oertel E, Manske E. Using micro spheres as reference artifacts for the in-situ characterization of tactile 3D micro probes along the probing sphere’s equator. tm - Technisches Messen. 2024;91(5):244-254. doi:10.1515/teme-2023-0164
BibTeX
@article{Oertel2024,
author = {Oertel, Erik and Manske, Eberhard},
title = {Using micro spheres as reference artifacts for the in-situ characterization of tactile 3D micro probes along the probing sphere’s equator},
journal = {tm - Technisches Messen},
year = {2024},
volume = {91},
number = {5},
pages = {244--254},
doi = {10.1515/teme-2023-0164}
}
EUVIMED