Pull-off characteristics of double-shanked compared to single-shanked ligation clips: an animal study
Innovative Surgical Sciences · 2016
DOI: 10.1515/iss-2016-0003 ↗Kurz gefasst
Fachartikel in der Zeitschrift Innovative Surgical Sciences, 2016. Verfasst von Martin Schenk und 5 weiteren Autor:innen. Beteiligte Einrichtung: Universitätsklinikum Tübingen, Klinik für Allgemeine, Viszeral- und Transplantationschirurgie, Institut für Experimentelle Chirurgie, Tübingen, Germa…. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
The use of surgical ligation clips is considered as the gold standard for the closure of vessels, particularly in laparoscopic surgery. The safety of clips is mainly achieved by the deep indentation of the metal bar with a high retention force. A novel double-shanked (DS) titanium clip was compared to two single-shanked clips with respect to axial and radial pull-off forces.
Methods
In a porcine model (8 animals, 51±1 kg), arteries were prepared immediately after euthanisation, assigned to either a medium (2–4 mm; n=120) or a medium-large (3.5–7 mm; n=120) clip size group, and clipped with the appropriate clip size. After dissection, axial and radial pull-off forces were measured.
Results
The axial pull-off force of the DS-Clip was higher than one single-shanked clip and comparable to the other single-shanked clip, and overall was linearly correlated to the cross-sectional area of the clip. The radial pull-off force of the DS-Clip was significantly higher than both single-shanked clips and, for the single-shanked clips, was correlated to the total clip thickness. The variation of radial pull-off force was lower for the DS-Clip due to a defined catch in the clip applier.
Conclusions
The radial pull-off force was lower than the axial pull-off force in total and therefore appears to be the critical point of dislocation. Due to the higher total holding mass, the DS-Clip provided a clear advantage in this regard and might therefore decrease the dislocation rate. The catch in the applier increases the reproducibility in clip placement.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Universitätsklinikum Tübingen, Klinik für Allgemeine, Viszeral- und Transplantationschirurgie, Institut für Experimentelle Chirurgie, Tübingen, Germany.
- Caritas Krankenhaus, Klinik für Allgemeine, Visceral- und Gefäßchirurgie, Bad Mergentheim, Germany.
Zitieren
APA
Schenk, M., Müller, G., Greiner, T. O., Fahrner, C., Königsrainer, A., & Thiel, C. (2016). Pull-off characteristics of double-shanked compared to single-shanked ligation clips: an animal study. Innovative Surgical Sciences. https://doi.org/10.1515/iss-2016-0003
Vancouver
Schenk M, Müller G, Greiner TO, Fahrner C, Königsrainer A, Thiel C. Pull-off characteristics of double-shanked compared to single-shanked ligation clips: an animal study. Innovative Surgical Sciences. 2016. doi:10.1515/iss-2016-0003
BibTeX
@article{Schenk2016,
author = {Schenk, Martin and Müller, Gottfried and Greiner, Tim Oliver and Fahrner, Christine and Königsrainer, Alfred and Thiel, Christian},
title = {Pull-off characteristics of double-shanked compared to single-shanked ligation clips: an animal study},
journal = {Innovative Surgical Sciences},
year = {2016},
doi = {10.1515/iss-2016-0003}
}
EUVIMED