Pre-analytical interference in point-of-care troponin T testing: a case series
Diagnosis · 2026
DOI: 10.1515/dx-2025-0104 ↗Kurz gefasst
Fallbericht in der Zeitschrift Diagnosis, 2026. Verfasst von Lekha Priyadharshini Kamarajan und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biochemistry, All India Institute of Medical Science Patna, Patna, India. Hauptthemen laut MEDLINE: Troponin T, Acute Coronary Syndrome, Point-of-Care Testing, Point-of-Care Systems. Die Arbeit nennt 10 Literaturangaben.
Abstract (Originalsprache)
Objectives
To examine the impact of haematological and biochemical abnormalities on the failure of point-of-care troponin T (POCT) testing in patients with suspected acute coronary syndrome (ACS). Case presentation Five patients underwent Roche Troponin T Card Test (lateral flow immunoassay) in an emergency setting. Despite correct sampling and procedural adherence, no valid results were obtained resulting in an aborted test. Laboratory analysis revealed severe anaemia, polycythaemia, leucocytosis, thrombocytopenia, and hepatic and renal dysfunction across cases. After stabilization, repeat POCT yielded valid results in all survivors, correlating with normalized haematological parameters.
Conclusions
Pre-analytical factors such as extreme haematocrit, leucocytosis, and biochemical derangements can cause POCT failure. Pre-testing screening and guideline updates are essential to optimize POCT reliability in acute care.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Biochemistry, All India Institute of Medical Science Patna, Patna, India.
- Department of Cardiology, All India Institute of Medical Science Patna, Patna, India.
- Department of Biochemistry, All India Institute of Medical Science Patna, Patna, Bihar, 801507, India.
ORCID-Profile: Mala Mahto
Zitieren
APA
Kamarajan, L. P., Tripathi, P., Kumar, S., Bhambhani, A., & Mahto, M. (2026). Pre-analytical interference in point-of-care troponin T testing: a case series. Diagnosis. https://doi.org/10.1515/dx-2025-0104
Vancouver
Kamarajan LP, Tripathi P, Kumar S, Bhambhani A, Mahto M. Pre-analytical interference in point-of-care troponin T testing: a case series. Diagnosis. 2026. doi:10.1515/dx-2025-0104
BibTeX
@article{Kamarajan2026,
author = {Kamarajan, Lekha Priyadharshini and Tripathi, Priyanshu and Kumar, Sushil and Bhambhani, Anupam and Mahto, Mala},
title = {Pre-analytical interference in point-of-care troponin T testing: a case series},
journal = {Diagnosis},
year = {2026},
doi = {10.1515/dx-2025-0104}
}
EUVIMED