Prospective real-time implementation and verification of patient-based real-time quality control using average of normals in a tertiary hospital laboratory
Clinical Chemistry and Laboratory Medicine (CCLM) · 25.11.2026 · 64 (12) · 2794-2803
DOI: 10.1515/cclm-2026-0201 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Chemistry and Laboratory Medicine (CCLM), online erschienen am 17.07.2026. Verfasst von Se-Eun Koo und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine and Genetics, Samsung Medical Center, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Tertiary Care Centers, Laboratories, Hospital. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 17 Literaturangaben.
Abstract (Originalsprache)
Objectives
Practical implementation of patient-based real-time quality control (PBRTQC) requires analyte- and institution-specific optimization. This study evaluated the operational feasibility of average-of-normals (AoN)-based PBRTQC through a structured selection process and prospective event-based validation.
Methods
PBRTQC monitoring was performed using the Average of Normals module in navify ® Lab Operations (Roche Diagnostics, Mannheim, Germany), which implements a simple moving average (SMA) algorithm. Candidate analytes were selected using retrospective patient results generated on cobas 8,000 series analyzers (Roche Diagnostics, Mannheim, Germany). Analyte selection was guided by a predefined, institution-specific algorithm incorporating distributional stability, biological variability, and operational characteristics. Visit-type stratification was considered when the absolute difference between the overall CV and the mean visit-type-specific CV exceeded 0.5 percentage points. PBRTQC parameters were optimized using a retrospective training dataset and subsequently verified in an independent dataset. Prospective validation was conducted using an event-based framework incorporating patient sample re-measurement and contemporaneous IQC evaluation.
Results
Based on the selection algorithm, creatinine, sodium, chloride, potassium, calcium, and free thyroxine were selected, with α-fetoprotein, aspartate aminotransferase, and alanine aminotransferase included for comparison. Across 14 PBRTQC parameter sets, 217,372 AoN points generated 1,599 alarms consolidated into 300 events. Calcium, chloride, creatinine (outpatient), and sodium showed high proportions of meaningful events (38–82 % of total events), whereas AFP and potassium predominantly generated non-confirmatory alarms.
Conclusions
AoN-based PBRTQC can complement conventional IQC for selected analytes. Its performance is strongly dependent on analyte characteristics and institution-specific patient result distributions, underscoring the need for ongoing monitoring and periodic re-evaluation to ensure sustained performance in routine laboratory practice.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Laboratory Medicine and Genetics, Samsung Medical Center, Seoul, Republic of Korea.
- Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
- Roche Diagnostics, Seoul, Republic of Korea.
- Department of Medical Device Management and Research, SAIHST, Sungkyunkwan University, Seoul, Republic of Korea.
Zitieren
APA
Koo, S. E., Shim, S., Park, Y. H., Kang, E., & Park, H. D. (2026). Prospective real-time implementation and verification of patient-based real-time quality control using average of normals in a tertiary hospital laboratory. Clinical Chemistry and Laboratory Medicine (CCLM), 64(12), 2794-2803. https://doi.org/10.1515/cclm-2026-0201
Vancouver
Koo SE, Shim S, Park YH, Kang E, Park HD. Prospective real-time implementation and verification of patient-based real-time quality control using average of normals in a tertiary hospital laboratory. Clinical Chemistry and Laboratory Medicine (CCLM). 2026;64(12):2794-2803. doi:10.1515/cclm-2026-0201
BibTeX
@article{Koo2026,
author = {Koo, Se-Eun and Shim, Sangwook and Park, Yeon-Hee and Kang, Eunji and Park, Hyung-Doo},
title = {Prospective real-time implementation and verification of patient-based real-time quality control using average of normals in a tertiary hospital laboratory},
journal = {Clinical Chemistry and Laboratory Medicine (CCLM)},
year = {2026},
volume = {64},
number = {12},
pages = {2794--2803},
doi = {10.1515/cclm-2026-0201}
}
EUVIMED