Prospective real-time implementation and verification of patient-based real-time quality control using average of normals in a tertiary hospital laboratory

Se-Eun Koo, Sangwook Shim, Yeon-Hee Park, Eunji Kang, Hyung-Doo Park

Clinical Chemistry and Laboratory Medicine (CCLM) · 25.11.2026 · 64 (12) · 2794-2803

DOI: 10.1515/cclm-2026-0201 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Chemistry and Laboratory Medicine (CCLM), online erschienen am 17.07.2026. Verfasst von Se-Eun Koo und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine and Genetics, Samsung Medical Center, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Tertiary Care Centers, Laboratories, Hospital. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 17 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Practical implementation of patient-based real-time quality control (PBRTQC) requires analyte- and institution-specific optimization. This study evaluated the operational feasibility of average-of-normals (AoN)-based PBRTQC through a structured selection process and prospective event-based validation.

Methods

PBRTQC monitoring was performed using the Average of Normals module in navify ® Lab Operations (Roche Diagnostics, Mannheim, Germany), which implements a simple moving average (SMA) algorithm. Candidate analytes were selected using retrospective patient results generated on cobas 8,000 series analyzers (Roche Diagnostics, Mannheim, Germany). Analyte selection was guided by a predefined, institution-specific algorithm incorporating distributional stability, biological variability, and operational characteristics. Visit-type stratification was considered when the absolute difference between the overall CV and the mean visit-type-specific CV exceeded 0.5 percentage points. PBRTQC parameters were optimized using a retrospective training dataset and subsequently verified in an independent dataset. Prospective validation was conducted using an event-based framework incorporating patient sample re-measurement and contemporaneous IQC evaluation.

Results

Based on the selection algorithm, creatinine, sodium, chloride, potassium, calcium, and free thyroxine were selected, with α-fetoprotein, aspartate aminotransferase, and alanine aminotransferase included for comparison. Across 14 PBRTQC parameter sets, 217,372 AoN points generated 1,599 alarms consolidated into 300 events. Calcium, chloride, creatinine (outpatient), and sodium showed high proportions of meaningful events (38–82 % of total events), whereas AFP and potassium predominantly generated non-confirmatory alarms.

Conclusions

AoN-based PBRTQC can complement conventional IQC for selected analytes. Its performance is strongly dependent on analyte characteristics and institution-specific patient result distributions, underscoring the need for ongoing monitoring and periodic re-evaluation to ensure sustained performance in routine laboratory practice.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tertiary Care CentersLaboratories, HospitalHumansQuality ControlAlgorithmsProspective StudiesLaboratories, Clinical

Stichwörter der Autor:innen

average of normalsclinical laboratory techniquespatient-based real-time quality controlprospective studiesquality control

Autor:innen und Institutionen

  • Department of Laboratory Medicine and Genetics, Samsung Medical Center, Seoul, Republic of Korea.
  • Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Roche Diagnostics, Seoul, Republic of Korea.
  • Department of Medical Device Management and Research, SAIHST, Sungkyunkwan University, Seoul, Republic of Korea.

Zitieren

APA

Koo, S. E., Shim, S., Park, Y. H., Kang, E., & Park, H. D. (2026). Prospective real-time implementation and verification of patient-based real-time quality control using average of normals in a tertiary hospital laboratory. Clinical Chemistry and Laboratory Medicine (CCLM), 64(12), 2794-2803. https://doi.org/10.1515/cclm-2026-0201

Vancouver

Koo SE, Shim S, Park YH, Kang E, Park HD. Prospective real-time implementation and verification of patient-based real-time quality control using average of normals in a tertiary hospital laboratory. Clinical Chemistry and Laboratory Medicine (CCLM). 2026;64(12):2794-2803. doi:10.1515/cclm-2026-0201

BibTeX
@article{Koo2026,
  author = {Koo, Se-Eun and Shim, Sangwook and Park, Yeon-Hee and Kang, Eunji and Park, Hyung-Doo},
  title = {Prospective real-time implementation and verification of patient-based real-time quality control using average of normals in a tertiary hospital laboratory},
  journal = {Clinical Chemistry and Laboratory Medicine (CCLM)},
  year = {2026},
  volume = {64},
  number = {12},
  pages = {2794--2803},
  doi = {10.1515/cclm-2026-0201}
}
Kostenlos starten →