EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Diagnostic sample transport via pneumatic tube systems: data logger and their algorithms are sensitive to transport effects

Jana Ninnemann, Stephanie Zylla, Thomas Streichert iD, Benjamin Otto, Mattis Haenel, Matthias Nauck, Astrid Petersmann

Clinical Chemistry and Laboratory Medicine (CCLM) · 2024

DOI: 10.1515/cclm-2023-0632 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Chemistry and Laboratory Medicine (CCLM), 2024. Verfasst von Jana Ninnemann und 6 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Clinical Chemistry and Laboratory Medicine, University Medicine Greifswald, Greifswald, Germany. Hauptthemen laut MEDLINE: Hemolysis, Blood Specimen Collection. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Many hospitals use pneumatic tube systems (PTS) for transport of diagnostic samples. Continuous monitoring of PTS and evaluation prior to clinical use is recommended. Data loggers with specifically developed algorithms have been suggested as an additional tool in PTS evaluation. We compared two different data loggers.

Methods

Transport types – courier, conventional (cPTS) and innovative PTS (iPTS) – were monitored using two data loggers (MSR145 ® logger, CiK Solutions GmbH, Karlsruhe, Germany, and a prototype developed at the University Medicine Greifswald). Data loggers differ in algorithm, recording frequencies and limit of acceleration detection. Samples from apparently healthy volunteers were split among the transport types and results for 37 laboratory measurands were compared.

Results

For each logger specific arbitrary units were calculated. Area-under-the-curve (AUC)-values (MSR145 ® ) were lowest for courier and highest for iPTS and increased with increasing recording frequencies. Stress (St)-values (prototype logger) were obtained in kmsu (1,000*mechanical stress unit) and were highest for iPTS as well. Statistical differences between laboratory measurement results of transport types were observed for three measurands sensitive for hemolysis.

Conclusions

The statistical, but not clinical, differences in the results for hemolysis sensitive measurands may be regarded as an early sign of preanalytical impairment. Both data loggers record this important interval of beginning mechanical stress with a high resolution indicating their potential to facilitate early detection of preanalytical impairment. Further studies should identify suitable recording frequencies. Currently, evaluation and monitoring of diagnostic sample transport should not only rely on data loggers but also include diagnostic samples.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HemolysisBlood Specimen CollectionHumansStress, MechanicalGermany

Stichwörter der Autor:innen

algorithmdata loggerdiagnostic sample transport monitoringpneumatic tube systempreanalytical phase

Autor:innen und Institutionen

  • Institute of Clinical Chemistry and Laboratory Medicine, University Medicine Greifswald, Greifswald, Germany.
  • DZHK (German Centre for Cardiovascular Research), Partner Site Greifswald, University Medicine Greifswald, Greifswald, Germany.
  • Institute of Clinical Chemistry, Faculty of Medicine and University Hospital, University Hospital Cologne, Cologne, Germany.
  • Department of Internal Medicine, University Medical Center, Hamburg-Eppendorf, Hamburg, Germany.
  • Max Planck Institute of Plasma Physics, Sub-institute Greifswald, Greifswald, Germany.
  • Institute of Clinical Chemistry and Laboratory Medicine, University Medicine Oldenburg, Oldenburg, Germany.

ORCID-Profile: Thomas Streichert

Zitieren

APA

Ninnemann, J., Zylla, S., Streichert, T., Otto, B., Haenel, M., Nauck, M., & Petersmann, A. (2024). Diagnostic sample transport via pneumatic tube systems: data logger and their algorithms are sensitive to transport effects. Clinical Chemistry and Laboratory Medicine (CCLM). https://doi.org/10.1515/cclm-2023-0632

Vancouver

Ninnemann J, Zylla S, Streichert T, Otto B, Haenel M, Nauck M, et al. Diagnostic sample transport via pneumatic tube systems: data logger and their algorithms are sensitive to transport effects. Clinical Chemistry and Laboratory Medicine (CCLM). 2024. doi:10.1515/cclm-2023-0632

BibTeX
@article{Ninnemann2024,
  author = {Ninnemann, Jana and Zylla, Stephanie and Streichert, Thomas and Otto, Benjamin and Haenel, Mattis and Nauck, Matthias and Petersmann, Astrid},
  title = {Diagnostic sample transport via pneumatic tube systems: data logger and their algorithms are sensitive to transport effects},
  journal = {Clinical Chemistry and Laboratory Medicine (CCLM)},
  year = {2024},
  doi = {10.1515/cclm-2023-0632}
}
Kostenlos starten →