An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure for the quantification of topiramate in human serum and plasma
Clinical Chemistry and Laboratory Medicine (CCLM) · 2023
DOI: 10.1515/cclm-2022-1273 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Chemistry and Laboratory Medicine (CCLM), 2023. Verfasst von Linda Salzmann und 11 weiteren Autor:innen. Beteiligte Einrichtung: Dr. Risch Ostschweiz AG, Buchs, Switzerland. Hauptthemen laut MEDLINE: Tandem Mass Spectrometry, Plasma. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Roche Diagnostics GmbH.
Abstract (Originalsprache)
Objectives
Topiramate is an antiepileptic drug (AED) used for the monotherapy or adjunctive treatment of epilepsy and for the prophylaxis of migraine. It has several pharmacodynamic properties that contribute to both its clinically useful properties and observed adverse effects. Accurate measurement of its concentration is therefore essential for dose adjustment/optimisation of AED therapy. Our aim was to develop and validate a novel reference measurement procedure (RMP) for the quantification of topiramate in human serum and plasma.
Methods
An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) method in combination with a protein-precipitation-based sample preparation allows for quantification of topiramate in human serum and plasma. To assure traceability to SI units, quantitative nuclear magnetic resonance (qNMR) was applied to characterize the reference material used as primary calibrator for this RMP. Matrix effects were determined by performing a post-column infusion experiment and comparing standard line slopes. Accuracy and precision was evaluated performing an extensive five day precision experiment and measurement uncertainty was evaluated according Guide to the Expression of Uncertainty in Measurement (GUM).
Results
The method enabled topiramate quantification within the range of 1.20–36.0 μg/mL without interference from structurally related compounds and no evidence of a matrix effect. Intermediate precision was ≤3.2 % and repeatability was 1.4–2.5 % across all concentration levels. The relative mean bias was −0.3 to 3.5 %. Expanded measurement uncertainties for target value assignment (n=6) were found to be ≤2.9 % (k=2) independent of the concentration level and the nature of the sample.
Conclusions
In human serum and plasma, the RMP demonstrated high analytical performance for topiramate quantification and fulfilled the requirements on measurement uncertainty. Traceability to SI units was established by qNMR content determination of the topiramate, which was used for direct calibration of the RMP. This RMP is, therefore, fit for purpose for routine assay standardization and clinical sample evaluation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Dr. Risch Ostschweiz AG, Buchs, Switzerland.
- Roche Diagnostics GmbH, Penzberg, Germany.
- Chrestos Concept GmbH & Co. KG, Essen, Germany.
ORCID-Profile: Lorenz Risch
Förderung
- Roche Diagnostics GmbH
Zitieren
APA
Salzmann, L., Spescha, T., Singh, N., Kobel, A., Fischer, V., Schierscher, T., Bauland, F., Geistanger, A., Risch, L., Geletneky, C., Seger, C., & Taibon, J. (2023). An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure for the quantification of topiramate in human serum and plasma. Clinical Chemistry and Laboratory Medicine (CCLM). https://doi.org/10.1515/cclm-2022-1273
Vancouver
Salzmann L, Spescha T, Singh N, Kobel A, Fischer V, Schierscher T, et al. An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure for the quantification of topiramate in human serum and plasma. Clinical Chemistry and Laboratory Medicine (CCLM). 2023. doi:10.1515/cclm-2022-1273
BibTeX
@article{Salzmann2023,
author = {Salzmann, Linda and Spescha, Tino and Singh, Neeraj and Kobel, Anja and Fischer, Vanessa and Schierscher, Tobias and Bauland, Friederike and Geistanger, Andrea and Risch, Lorenz and Geletneky, Christian and Seger, Christoph and Taibon, Judith},
title = {An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure for the quantification of topiramate in human serum and plasma},
journal = {Clinical Chemistry and Laboratory Medicine (CCLM)},
year = {2023},
doi = {10.1515/cclm-2022-1273}
}
EUVIMED