Establishment of the isotope dilution-gas chromatography-tandem mass spectrometry (ID-GC/MS) method for accurate quantification of phenylalanine and its application to value assignment of external quality assessment samples

Zhizheng Li, Zhonggan Jin, Ming Zong

Clinica Chimica Acta · 2025

DOI: 10.1016/j.cca.2025.120262 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinica Chimica Acta, 2025. Verfasst von Zhizheng Li und 2 weiteren Autor:innen. Beteiligte Einrichtung: School of Medical Technology, Shanghai University of Medicine & Health Sciences, Shanghai 201318, China. Hauptthemen laut MEDLINE: Phenylalanine, Gas Chromatography-Mass Spectrometry, Tandem Mass Spectrometry. Die Arbeit nennt 30 Literaturangaben. Gefördert von: Shanghai Municipal Health Commission, Shanghai Municipal Health and Family Planning Commission, Shanghai Pudong New Area Health Commission.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVES: Phenylalanine (Phe) was an essential amino acid that was crucial for diagnosing phenylketonuria (PKU) and managing their blood Phe levels. Currently, there was no JCTLM-recognized reference method for measuring Phe internationally. To standardize the measurement of serum Phe, our primary goal was to establish an accurate quantitative method based on isotope dilution combined with gas chromatography-mass spectrometry (ID-GC/MS) for the quantitative detection of Phe in human serum, and its performance was thoroughly validated. METHODS: Serum samples were pre-treated using acetonitrile protein precipitation. After derivatization with N-(Methyl)-N-(trimethylsilyl)trifluoroacetamide (MSTFA), serum Phe content was quantitatively detected using the ID-GC/MS method. Mass spectrometry analysis was conducted using the SIM monitoring mode. The analytical performance and application of the new method were investigated. External quality assessment (EQA) samples were tested using the established method, and their accuracy was compared with clinical laboratory results. RESULTS: The total detection time for Phe using the established method was 7 min. This method was capable of quantifying Phe within the range of 7.39 to 1188.81 μmol/L, with a limit of quantitation of 3.03 μmol/L. The intra-assay and inter-assay coefficients of variation (CV) were both ≤2.5 %, and the spike recovery rate ranged from 97.64 % to 100.72 %, and a relative expanded uncertainty was ≤3.0 % (k = 2). The method was free from interference, matrix effect and carry-over. Moreover, the ID-GC/MS measurement procedure was successfully applied to measure Phe in serum samples and to assign value of EQA samples. CONCLUSIONS: This study successfully established an accurate quantitative method for detecting serum Phe concentrations using ID-GC/MS technology and it could support EQA management for Phe testing, providing traceable target values for samples.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PhenylalanineGas Chromatography-Mass SpectrometryTandem Mass SpectrometryHumansQuality ControlIndicator Dilution TechniquesPhenylketonuriasIsotopes

Stichwörter der Autor:innen

Accurate quantitative methodAmino acidsIsotope dilution-gas chromatography-tandem mass spectrometryLaboratory medicinePhenylalanine

Substanzen

PhenylalanineIsotopes

Autor:innen und Institutionen

  • School of Medical Technology, Shanghai University of Medicine & Health Sciences, Shanghai 201318, China.
  • Shanghai Center for Clinical Laboratory, Shanghai 200126, China. Electronic address: jinzhonggan@sccl.org.cn.
  • Department of Clinical Laboratory, Shanghai East Hospital, Tongji University School of Medicine, Shanghai 200120, China. Electronic address: zongming@tongji.edu.cn.

Förderung

  • Shanghai Municipal Health Commission
  • Shanghai Municipal Health and Family Planning Commission
  • Shanghai Pudong New Area Health Commission
  • Shanghai University of Medicine and Health Sciences
  • Shanghai Chest Hospital

Zitieren

APA

Li, Z., Jin, Z., & Zong, M. (2025). Establishment of the isotope dilution-gas chromatography-tandem mass spectrometry (ID-GC/MS) method for accurate quantification of phenylalanine and its application to value assignment of external quality assessment samples. Clinica Chimica Acta. https://doi.org/10.1016/j.cca.2025.120262

Vancouver

Li Z, Jin Z, Zong M. Establishment of the isotope dilution-gas chromatography-tandem mass spectrometry (ID-GC/MS) method for accurate quantification of phenylalanine and its application to value assignment of external quality assessment samples. Clinica Chimica Acta. 2025. doi:10.1016/j.cca.2025.120262

BibTeX
@article{Li2025,
  author = {Li, Zhizheng and Jin, Zhonggan and Zong, Ming},
  title = {Establishment of the isotope dilution-gas chromatography-tandem mass spectrometry (ID-GC/MS) method for accurate quantification of phenylalanine and its application to value assignment of external quality assessment samples},
  journal = {Clinica Chimica Acta},
  year = {2025},
  doi = {10.1016/j.cca.2025.120262}
}
Kostenlos starten →