Noise reduction and QRS detection in ECG signal using EEMD with modified sigmoid thresholding
Biomedical Engineering / Biomedizinische Technik · 2024
DOI: 10.1515/bmt-2022-0450 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biomedical Engineering / Biomedizinische Technik, 2024. Verfasst von Ouahiba Mohguen. Beteiligte Einrichtung: Department of Electronics, LIS Laboratory University Ferhat Abbas Setif 1, Setif, Algeria. Hauptthemen laut MEDLINE: Signal Processing, Computer-Assisted, Algorithms. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 43 Literaturangaben.
Abstract (Originalsprache)
Objectives
Novel noise reduction and QRS detection algorithms in Electrocardiogram (ECG) signal based on Empirical Mode Decomposition (EMD), Ensemble Empirical Mode Decomposition (EEMD) and the Modified Sigmoid Thresholding Function (MSTF) are proposed in this paper.
Methods
EMD and EEMD algorithms are used to decompose the noisy ECG signal into series of Intrinsic Mode Functions (IMFs). Then, these IMFs are thresholded by the MSTF for reduction of noises and preservation of QRS complexes. After that, the thresholded IMFs are used to obtain the clean ECG signal. The characteristic points P, Q, R, S and T peaks are detected using peak detection algorithm.
Results
The proposed methods are validated through experiments on the MIT-BIH arrhythmia database and Additive White Gaussian Noise (AWGN) is added to the clean ECG signal at different input SNR ( SNR in ). Standard performance parameters output SNR ( SNR out ), mean square error ( MSE ), root mean square error ( RMSE ), SNR improvement ( SNR imp ) and percentage root mean square difference ( PRD ) are employed for evaluation of the efficacy of the proposed methods. The results showed that the proposed methods provide significant quantitative and qualitative improvements in denoising performance, compared with existing state-of-the-art methods such as wavelet denoising, conventional EMD (EMD-Conv), conventional EEMD (EEMD-Conv, Stockwell Transform (ST) and Complete EEMD with Adaptative Noise with hybrid interval thresholding and higher order statistic to select relevant modes (CEEMDAN-HIT).
Conclusions
A detail quantitative analysis demonstrate that for abnormal ECG records 207 m and 214 m at input SNR of −2 dB the SNR imp value is 12.22 and 11.58 dB respectively, which indicates that the proposed algorithm can be used as an effective tool for denoising of ECG signals.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electronics, LIS Laboratory University Ferhat Abbas Setif 1, Setif, Algeria.
Zitieren
APA
Mohguen, O. (2024). Noise reduction and QRS detection in ECG signal using EEMD with modified sigmoid thresholding. Biomedical Engineering / Biomedizinische Technik. https://doi.org/10.1515/bmt-2022-0450
Vancouver
Mohguen O. Noise reduction and QRS detection in ECG signal using EEMD with modified sigmoid thresholding. Biomedical Engineering / Biomedizinische Technik. 2024. doi:10.1515/bmt-2022-0450
BibTeX
@article{Mohguen2024,
author = {Mohguen, Ouahiba},
title = {Noise reduction and QRS detection in ECG signal using EEMD with modified sigmoid thresholding},
journal = {Biomedical Engineering / Biomedizinische Technik},
year = {2024},
doi = {10.1515/bmt-2022-0450}
}
EUVIMED