Simultaneous validation of wearable motion capture system for lower body applications: over single plane range of motion (ROM) and gait activities
Biomedical Engineering / Biomedizinische Technik · 27.06.2022 · 67 (3) · 185-199
DOI: 10.1515/bmt-2021-0429 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biomedical Engineering / Biomedizinische Technik, online erschienen am 16.05.2022. Verfasst von Senay Mihcin. Beteiligte Einrichtung: Mechanical Engineering Department, Izmir Institute of Technology (IZTECH), Urla, Izmir, Turkey. Hauptthemen laut MEDLINE: Gait, Wearable Electronic Devices. Die Arbeit wurde bisher 29-mal zitiert (Crossref) und nennt 39 Literaturangaben. Gefördert von: Türkiye Bilimsel ve Teknolojik Arastirma Kurumu.
Abstract (Originalsprache)
Extracting data from {Zhu, 2019 #5} daily life activities is important in biomechanical applications to define exact boundary conditions for the intended use-based applications. Although optoelectronic camera-marker based systems are used as gold standard tools for medical applications, due to line-of-sight problem, there is a need for wearable, affordable motion capture (MOCAP) systems. We investigate the potential use of a wearable inertial measurement unit (IMU) based-wearable MOCAP system for biomechanical applications. The in vitro proof of concept is provided for the full lower body consisting of hip, knee, and ankle joints via controlled single-plane anatomical range of motion (ROM) simulations using an electrical motor, while collecting data simultaneously via opto-electronic markers and IMU sensors. On 15 healthy volunteers the flexion-extension, abduction-adduction, internal-external rotation (ROM) values of hip and, the flexion – extension ROM values of the knee and ankle joints are calculated for both systems. The Bland-Altman graphs showed promising agreement both for in vitro and in vivo experiments. The maximum Root Mean Square Errors (RMSE) between the systems in vitro was 3.4° for hip and 5.9° for knee flexion motion in vivo , respectively. The gait data of the volunteers were assessed between the heel strike and toe off events to investigate the limits of agreement, calculating the population averages and standard deviation for both systems over the gait cycle. The maximum difference was for the ankle joint
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Mechanical Engineering Department, Izmir Institute of Technology (IZTECH), Urla, Izmir, Turkey.
ORCID-Profile: Senay Mihcin
Förderung
- Türkiye Bilimsel ve Teknolojik Arastirma Kurumu
Zitieren
APA
Mihcin, S. (2022). Simultaneous validation of wearable motion capture system for lower body applications: over single plane range of motion (ROM) and gait activities. Biomedical Engineering / Biomedizinische Technik, 67(3), 185-199. https://doi.org/10.1515/bmt-2021-0429
Vancouver
Mihcin S. Simultaneous validation of wearable motion capture system for lower body applications: over single plane range of motion (ROM) and gait activities. Biomedical Engineering / Biomedizinische Technik. 2022;67(3):185-199. doi:10.1515/bmt-2021-0429