EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Assessment of the MX3 hydration testing system in adults at rest and following exercise in the heat

Xiujing Zhao iD, Zachary J McKenna iD, Rosie I Perez, Shawn C Wierick, Becca M Allen, Brendon P McDermott iD

Physiological Reports · 2026-03 · 14 (6)

DOI: 10.14814/phy2.70757 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physiological Reports, online erschienen am 26.03.2026. Verfasst von Xiujing Zhao und 5 weiteren Autor:innen. Beteiligte Einrichtung: Heat & Hydration Optimization (H2O) Lab, University of Arkansas, Fayetteville, Arkansas, USA. Hauptthemen laut MEDLINE: Hot Temperature, Saliva, Exercise, Rest. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study examined the performance of MX3 hydration testing system for assessing hydration status at rest and following exercise. Twenty adults completed two matched exercise trials in warm‐humid conditions. Pre‐ and post‐exercise saliva osmolality (S osm ) were assessed using MX3. Reliability was assessed using mean difference (MD), absolute difference (ABD), coefficient of variation (CV), standard error of the measurement (SEM), intraclass correlation coefficient (ICC), and 95% limits of agreement (LoA). Diagnostic accuracy was assessed using a crosstab analysis relative to urine specific gravity (USG) and body mass loss (BML). Pre‐exercise, MX3 showed a MD of 4 mOsm/kg (95% LoA: −46, 54), ABD of 23 mOsm/kg, CV of 31%, SEM of 18 mOsm/kg, and ICC of 0.198. Post‐exercise, MX3 demonstrated a MD of 5 mOsm/kg (95% LoA: −78, 68), ABD of 28 mOsm/kg, CV of 29%, SEM of 26 mOsm/kg, and ICC of 0.391. Pre‐exercise, S osm demonstrated a sensitivity of 0.09, specificity of 0.82, and LR + = 0.51, LR − = 1.11 relative to USG >1.020. Post‐exercise, S osm showed a sensitivity of 0.57, specificity of 0.63, and LR + = 1.52, LR − = 0.69 relative to >2% BML. MX3 has low reliability and poor‐to‐weak predictive value in identifying hypohydration at rest and following exercise.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hot TemperatureSalivaExerciseRestDehydrationOrganism Hydration StatusHumansOsmolar ConcentrationMaleFemaleAdultYoung AdultReproducibility of Results

Stichwörter der Autor:innen

exerciseheathydrationsaliva osmolality

Autor:innen und Institutionen

  • Heat & Hydration Optimization (H2O) Lab, University of Arkansas, Fayetteville, Arkansas, USA.
  • Exercise Science Research Center, University of Arkansas, Fayetteville, Arkansas, USA.
  • Department of Health, Human Performance & Recreation, University of Arkansas, Fayetteville, Arkansas, USA.

ORCID-Profile: Xiujing Zhao, Zachary J McKenna, Brendon P McDermott

Zitieren

APA

Zhao, X., McKenna, Z. J., Perez, R. I., Wierick, S. C., Allen, B. M., & McDermott, B. P. (2026). Assessment of the MX3 hydration testing system in adults at rest and following exercise in the heat. Physiological Reports, 14(6). https://doi.org/10.14814/phy2.70757

Vancouver

Zhao X, McKenna ZJ, Perez RI, Wierick SC, Allen BM, McDermott BP. Assessment of the MX3 hydration testing system in adults at rest and following exercise in the heat. Physiological Reports. 2026;14(6). doi:10.14814/phy2.70757

BibTeX
@article{Zhao2026,
  author = {Zhao, Xiujing and McKenna, Zachary J and Perez, Rosie I and Wierick, Shawn C and Allen, Becca M and McDermott, Brendon P},
  title = {Assessment of the MX3 hydration testing system in adults at rest and following exercise in the heat},
  journal = {Physiological Reports},
  year = {2026},
  volume = {14},
  number = {6},
  doi = {10.14814/phy2.70757}
}
Kostenlos starten →