Screw Insertional Torque Measurement in Spine Surgery: Correlation With Bone Mineral Density and Hounsfield Unit
Neurospine · 31.12.2023 · 20 (4) · 1177-1185
DOI: 10.14245/ns.2346830.415 ↗Kurz gefasst
Fachartikel in der Zeitschrift Neurospine, erschienen am 31.12.2023. Verfasst von Byeong Ho Oh und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurosurgery, Chungbuk National University Hospital, Cheongju, Korea. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 40 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Objective
Achieving successful fusion during spine surgery is dependent on rigid pedicle screw fixation. To assess fixation strength, the insertional torque can be measured during intraoperative screw fixation. This study aimed to explore the technical feasibility of measuring the insertional torque of a pedicle screw, while investigating its relationship with bone density.
Methods
Thoraco-lumbar screw fixation fusion surgery was performed on 53 patients (mean age, 65.5 ± 9.8 years). The insertional torque of 284 screws was measured at the point passing through the pedicle using a calibrated torque wrench, with a specially designed connector to the spine screw system. The Hounsfield units (HU) value was determined by assessing the trabecular portion of the index vertebral body on sagittal computed tomography images. We analyzed the relationship between the measured insertional torque and the following bone strength parameters: bone mineral density (BMD) and HU of the vertebral body.
Results
The mean insertion torque was 105.55 ± 58.08 N∙cm and T-score value (BMD) was -1.14 ± 1.49. Mean HU value was 136.37 ± 57.59. Screw insertion torque was positively correlated with BMD and HU in whole patients. However, in cases of osteopenia, all variables showed very weak correlations with insertional torque. In patients with osteoporosis, there was no statistically significant correlation between BMD and torque strength; HU showed a significant correlation.
Conclusion
The insertional torque of screw fixation significantly correlated with bone density (BMD and HU). HU measurements showed greater clinical significance than did BMD values in patients with osteoporosis.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Neurosurgery, Chungbuk National University Hospital, Cheongju, Korea.
- Department of Neurosurgery, St. Vincent's Hospital, The Catholic University of Korea, Suwon, Korea.
- Department of Neurosurgery, Eunpyeong St. Mary's Hospital, The Catholic University of Korea, Seoul, Korea.
- Department of Neurological Surgery, Duke University Medical Center, Durham, NC, USA.
Zitieren
APA
Oh, B. H., Kim, J. Y., Lee, J. B., Hong, J. T., Sung, J. H., Than, K. D., Lee, H. J., & Kim, I. S. (2023). Screw Insertional Torque Measurement in Spine Surgery: Correlation With Bone Mineral Density and Hounsfield Unit. Neurospine, 20(4), 1177-1185. https://doi.org/10.14245/ns.2346830.415
Vancouver
Oh BH, Kim JY, Lee JB, Hong JT, Sung JH, Than KD, et al. Screw Insertional Torque Measurement in Spine Surgery: Correlation With Bone Mineral Density and Hounsfield Unit. Neurospine. 2023;20(4):1177-1185. doi:10.14245/ns.2346830.415
BibTeX
@article{Oh2023,
author = {Oh, Byeong Ho and Kim, Jee Yong and Lee, Jong Beom and Hong, Jae Taek and Sung, Jae Hoon and Than, Khoi D and Lee, Ho Jin and Kim, Il Sup},
title = {Screw Insertional Torque Measurement in Spine Surgery: Correlation With Bone Mineral Density and Hounsfield Unit},
journal = {Neurospine},
year = {2023},
volume = {20},
number = {4},
pages = {1177--1185},
doi = {10.14245/ns.2346830.415}
}
EUVIMED