EUVIMEDEUVIMED entdecken
Randomisierte kontrollierte StudieOpen Access · CC BY 4.0

Microbiological surveillance after automated reprocessing of gastrointestinal endoscopes in routine clinical practice: A prospective randomized microbiological surveillance study

Masayuki Kato iD, Hideki Hayashi, Nobuharu Hironaka, Masafumi Chiba, Takahiro Abe, Nana Shimamoto, Mamoru Ito, Masakuni Kobayashi, Kazuki Sumiyama

PLOS One · 2026

DOI: 10.1371/journal.pone.0360048 ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift PLOS One, 2026. Verfasst von Masayuki Kato und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Endoscopy, The Jikei University School of Medicine, Tokyo, Japan. Hauptthemen laut MEDLINE: Endoscopes, Gastrointestinal, Disinfection, Equipment Contamination, Equipment Reuse. Die Arbeit nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Kaigen Pharma Co., Ltd., IHI Logistics & Machinery Corporation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Microbiological surveillance is used to monitor the quality of gastrointestinal endoscope reprocessing, but prospective data comparing automated endoscope reprocessor systems under routine workflow conditions remain limited. We conducted an exploratory, single-center, prospective randomized surveillance study of 222 endoscope-use episodes involving patient-used forward-viewing upper and lower gastrointestinal endoscopes. Endoscope-use episodes were allocated 1:1:1 to strongly acidic electrolyzed water, ozonated water, or peracetic acid using an Excel-generated random-number procedure (n = 74 each). Assigned reprocessing was performed after standardized bedside pre-cleaning and manual cleaning. Post-reprocessing samples were obtained from predefined external and internal sites. Surveillance success was defined as ≤20 colony-forming units per sample at all sampled sites and absence of predefined indicator organisms. Success was achieved in 67/74 (90.5%) strongly acidic electrolyzed-water episodes, 71/74 (95.9%) ozonated-water episodes, and 71/74 (95.9%) peracetic-acid episodes. Risk differences in surveillance success versus peracetic acid were −5.4 percentage points (95% confidence interval, −14.6 to 3.2) and 0.0 percentage points (95% confidence interval, −7.7 to 7.7), respectively. In exploratory post hoc non-inferiority analyses using a −5 percentage-point margin, non-inferiority was not demonstrated for either functional-water system. A −10 percentage-point margin supported ozonated water but not strongly acidic electrolyzed water. Post hoc scope-level sensitivity analyses were directionally consistent with the primary analysis and highlighted the influence of repeated failures in an individual physical endoscope. These findings should be interpreted as exploratory microbiological quality-assurance data, not as evidence of clinical infection risk or comparative clinical effectiveness. Study registration: University Hospital Medical Information Network Clinical Trials Registry (UMIN-CTR), UMIN000061627 (registered May 19, 2026; enrollment began November 8, 2024).

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Endoscopes, GastrointestinalDisinfectionEquipment ContaminationEquipment ReuseHumansProspective StudiesPeracetic Acid

Substanzen

Peracetic Acid

Autor:innen und Institutionen

  • Department of Endoscopy, The Jikei University School of Medicine, Tokyo, Japan.
  • Kaigen Pharma Co., Ltd., Osaka, Japan.
  • IHI Logistics & Machinery Corporation, Tokyo, Japan.

ORCID-Profile: Masayuki Kato

Förderung

  • Kaigen Pharma Co., Ltd.
  • IHI Logistics & Machinery Corporation

Zitieren

APA

Kato, M., Hayashi, H., Hironaka, N., Chiba, M., Abe, T., Shimamoto, N., Ito, M., Kobayashi, M., & Sumiyama, K. (2026). Microbiological surveillance after automated reprocessing of gastrointestinal endoscopes in routine clinical practice: A prospective randomized microbiological surveillance study. PLOS One. https://doi.org/10.1371/journal.pone.0360048

Vancouver

Kato M, Hayashi H, Hironaka N, Chiba M, Abe T, Shimamoto N, et al. Microbiological surveillance after automated reprocessing of gastrointestinal endoscopes in routine clinical practice: A prospective randomized microbiological surveillance study. PLOS One. 2026. doi:10.1371/journal.pone.0360048

BibTeX
@article{Kato2026,
  author = {Kato, Masayuki and Hayashi, Hideki and Hironaka, Nobuharu and Chiba, Masafumi and Abe, Takahiro and Shimamoto, Nana and Ito, Mamoru and Kobayashi, Masakuni and Sumiyama, Kazuki},
  title = {Microbiological surveillance after automated reprocessing of gastrointestinal endoscopes in routine clinical practice: A prospective randomized microbiological surveillance study},
  journal = {PLOS One},
  year = {2026},
  doi = {10.1371/journal.pone.0360048}
}
Kostenlos starten →