Anatomical fit of the oneKNEE tibia design using statistical shape modeling
PLOS One · 31.07.2026 · 21 (7) · e0354876
DOI: 10.1371/journal.pone.0354876 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS One, online erschienen am 31.07.2026. Verfasst von Ingrid Dupraz und 7 weiteren Autor:innen. Beteiligte Einrichtung: Research and Development, Aesculap AG, Tuttlingen, Germany. Hauptthemen laut MEDLINE: Tibia, Prosthesis Design, Models, Statistical, Knee Prosthesis. Die Arbeit nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction and aim Modern tibial implants should be designed to achieve an optimal anatomical fit. The aim of this study was to evaluate how well a newly developed tibial implant design (oneKNEE ® – B. Braun Aesculap, Tuttlingen, Germany) matches patient anatomy compared to three existing implant systems, using Statistical Shape Models (SSM).
Method
SSMs for Caucasian and Asian populations were generated using CT scans of 120 Caucasian and 112 Asian osteoarthritic patients. Border anatomical variations were represented by individual anatomies from 14 Caucasian and 12 Asian patients. Tibial components were positioned in accordance with strict protocols established by senior knee surgeons (YM, NK). The anatomical fit was analyzed across all four implant systems in terms of antero-posterior and medio-lateral dimensions, tibial coverage and cortical bone support. Global tests for homogeneity and Dunnett’s tests were performed.
Results
Bony coverage for Asian and Caucasian populations ranged from 82.9% to 88.4% for long-established designs and from 85.0% to 91.2% for more recently introduced designs. Among all designs, the oneKNEE ® implant demonstrated superior bony coverage and cortical support, followed by Attune ® (Depuy-Synthes, Warsaw, IN, USA), Columbus ® (B. Braun Aesculap), and PFC Sigma ® (Depuy-Synthes). For average anatomies, oneKNEE ® showed significantly better bony coverage and cortical score compared to Columbus ® and PFC Sigma ® (P < 0.01). In less average anatomies, oneKNEE ® exhibited statistically superior bony coverage compared to PFC Sigma ® , although no significant difference was observed in cortical support.
Conclusion
Compared with established tibial implant designs, the new design demonstrated improved virtual anatomical fit and coverage metrics based on computational modeling. These findings reflect a theoretical design‑level improvement and require biomechanical and clinical validation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Research and Development, Aesculap AG, Tuttlingen, Germany.
- Department of Orthopaedic Surgery, Osaka Metropolitan University Graduate School of Medicine, Osaka, Japan.
- Orthopedic Department N 17, Vreden National Medical Research Center of Traumatology and Orthopedics of the Ministry of Health of Russian Federation, Saint Petersburg, Russian Federation.
- Department of Orthopaedic and Trauma Surgery, Musculoskeletal University Center Munich (MUM), Campus Grosshadern, LMU Munich, Munich, Germany.
ORCID-Profile: Ingrid Dupraz, Berna Richter
Zitieren
APA
Dupraz, I., Minoda, Y., Kornilov, N., Altermann, B., Hauff, P., Bollinger, A., Richter, B., & Grupp, T. M. (2026). Anatomical fit of the oneKNEE tibia design using statistical shape modeling. PLOS One, 21(7), e0354876. https://doi.org/10.1371/journal.pone.0354876
Vancouver
Dupraz I, Minoda Y, Kornilov N, Altermann B, Hauff P, Bollinger A, et al. Anatomical fit of the oneKNEE tibia design using statistical shape modeling. PLOS One. 2026;21(7):e0354876. doi:10.1371/journal.pone.0354876
BibTeX
@article{Dupraz2026,
author = {Dupraz, Ingrid and Minoda, Yukihide and Kornilov, Nikolai and Altermann, Brigitte and Hauff, Philipp and Bollinger, Arthur and Richter, Berna and Grupp, Thomas M},
title = {Anatomical fit of the oneKNEE tibia design using statistical shape modeling},
journal = {PLOS One},
year = {2026},
volume = {21},
number = {7},
pages = {e0354876},
doi = {10.1371/journal.pone.0354876}
}
EUVIMED