EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

ThermiQuant™ AquaStream: A portable instrument for quantitative colorimetric isothermal nucleic acid amplification reactions in paper and tube formats

Bibek Raut iD, Gopal Palla, Jiangshan Wang iD, Skyler Campbell, Virendra Kumar iD, Kyungyeon Ra, Mohit S Verma iD

PLOS One · 15.05.2026 · 21 (5) · e0348607

DOI: 10.1371/journal.pone.0348607 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS One, online erschienen am 15.05.2026. Verfasst von Bibek Raut und 6 weiteren Autor:innen. Beteiligte Einrichtung: Weldon School of Biomedical Engineering, Purdue University, West Lafayette, Indiana, United States of America. Hauptthemen laut MEDLINE: Colorimetry, Nucleic Acid Amplification Techniques, Molecular Diagnostic Techniques, SARS-CoV-2. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 52 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Center for Produce Safety, Agricultural Marketing Service, Animal and Plant Health Inspection Service.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Microfluidic paper-based analytical devices (µPADs) are an attractive format for colorimetric nucleic acid amplification tests (NAATs) because they enable low-cost, portable diagnostics in resource-limited settings. However, researchers often optimize colorimetric assays in liquid reactions in tubes before translating them to µPADs. Since both formats require separate instruments for incubation and real-time sensing, direct comparison of reactions between the two formats is difficult. To address these cross-platform limitations, we developed ThermiQuant™ AquaStream, a portable benchtop device (15 × 20 × 16 cm, ~ 5 kg; cost: USD 327) that supports seamless colorimetric loop-mediated isothermal amplification (LAMP) reactions in both µPADs and tubes under a common workflow. The system enables real-time reaction tracking (every 30 seconds) through onboard image processing, precise isothermal control (±0.5 °C) using a repurposed consumer-grade sous-vide heater, and medium-throughput (24 tubes or 42 µPADs). Testing with synthetic SARS-CoV-2 orf7ab DNA fragments demonstrated a limit of detection corresponding to a 95% probability of detection (LOD95) of 110 copies per reaction in tube (22 copies/µL) and 39 copies per reaction in µPADs (5 copies/µL), estimated using probit regression. In both formats the limit of quantification (LOQ), defined as the lowest concentration yielding a coefficient of variation (CV) of quantification time (Tq) ≤ 10%, was 250 copies/reaction resulting in a strong linear (R 2 = 0.98 & R 2 = 0.96 respectively for tube and µPADs) standard calibration curves. ThermiQuant™ AquaStream provides an affordable and versatile benchtop platform capable of supporting both tube- and µPAD-based colorimetric LAMP assays, serving as a proof-of-concept research tool for assay development and molecular diagnostics in One Health settings such as clinics, farms, and field environments.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ColorimetryNucleic Acid Amplification TechniquesMolecular Diagnostic TechniquesSARS-CoV-2Rapid Diagnostic TestsPaperHumans

Autor:innen und Institutionen

  • Weldon School of Biomedical Engineering, Purdue University, West Lafayette, Indiana, United States of America.
  • Department of Agricultural and Biological Engineering, Purdue University, West Lafayette, Indiana, United States of America.
  • Department of Biomedical Engineering, Duke University, Durham, North Carolina, United States of America.
  • Purdue Institute of Inflammation, Immunology and Infectious Disease, Purdue University, West Lafayette, Indiana, United States of America.
  • Purdue Institute for a Sustainable Future, Purdue University, West Lafayette, Indiana, United States of America.
  • Bindley Bioscience Center, Purdue University, West Lafayette, Indiana, United States of America.

ORCID-Profile: Bibek Raut, Jiangshan Wang, Virendra Kumar, Mohit S Verma

Förderung

  • Center for Produce Safety
  • Agricultural Marketing Service
  • Animal and Plant Health Inspection Service

Zitieren

APA

Raut, B., Palla, G., Wang, J., Campbell, S., Kumar, V., Ra, K., & Verma, M. S. (2026). ThermiQuant™ AquaStream: A portable instrument for quantitative colorimetric isothermal nucleic acid amplification reactions in paper and tube formats. PLOS One, 21(5), e0348607. https://doi.org/10.1371/journal.pone.0348607

Vancouver

Raut B, Palla G, Wang J, Campbell S, Kumar V, Ra K, et al. ThermiQuant™ AquaStream: A portable instrument for quantitative colorimetric isothermal nucleic acid amplification reactions in paper and tube formats. PLOS One. 2026;21(5):e0348607. doi:10.1371/journal.pone.0348607

BibTeX
@article{Raut2026,
  author = {Raut, Bibek and Palla, Gopal and Wang, Jiangshan and Campbell, Skyler and Kumar, Virendra and Ra, Kyungyeon and Verma, Mohit S},
  title = {ThermiQuant™ AquaStream: A portable instrument for quantitative colorimetric isothermal nucleic acid amplification reactions in paper and tube formats},
  journal = {PLOS One},
  year = {2026},
  volume = {21},
  number = {5},
  pages = {e0348607},
  doi = {10.1371/journal.pone.0348607}
}