Development of a Loop-Mediated Isothermal Amplification (LAMP) for the screening of Candida auris
PLOS One · 24.04.2026 · 21 (4) · e0348003
DOI: 10.1371/journal.pone.0348003 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS One, online erschienen am 24.04.2026. Verfasst von Woong Sik Jang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Emergency Medicine, College of Medicine, Korea University Guro Hospital, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Nucleic Acid Amplification Techniques, Candida auris, Candidiasis, Molecular Diagnostic Techniques. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Korea University Guro Hospital, Korea Health Industry Development Institute.
Abstract (Originalsprache)
Background
Candida auris is an emerging multidrug-resistant yeast associated with invasive infections, healthcare-associated outbreaks, and high mortality, and is often misidentified by conventional diagnostic methods. Rapid, accurate, and scalable screening tools are essential for effective infection control, particularly in high-risk settings. Materials and methods We developed a multiplex loop-mediated isothermal amplification (LAMP) assay that combines a broad-range Candida Pan target with a C. auris –specific target in a single isothermal reaction. Assay conditions were optimized for primer ratio and temperature, and analytical sensitivity was evaluated using serial dilutions of culture-derived C. albicans and C. auris DNA, as well as contrived specimens consisting of urine, swab, and whole-blood matrices. Clinical performance was assessed using 35 Candida -positive clinical specimens (blood, urine, ear swabs) and 94 non-infectious controls. Results were compared with Candida Pan qPCR and C. auris qPCR. Cross-reactivity was tested against common bacterial isolates.
Results
Under optimized conditions (1:1 primer ratio, 64 °C), the assay allowed species-level discrimination, with C. auris positive for both Pan and auris channels and C. albicans positive only for the Pan channel. The C. auris -specific LAMP probe detected approximately 10²–10³ cells/mL in culture-derived and contrived specimens, showing a 1–2 log improvement over C. auris qPCR (10⁴–10⁵ cells/mL), while the Pan LAMP channel detected C. auris at around 10⁵ cells/mL. In clinical specimens, Pan LAMP detected Candida spp. in 34/35 cases (97.14%) versus 32/35 (91.14%) for Pan qPCR. All C. auris –positive specimens (9/9) were detected by the multiplex LAMP assay, compared with 6/9 (66.7%) by Pan qPCR. All 94 non-infectious controls and all bacterial isolates tested negative, indicating 100% clinical specificity and absence of cross-reactivity.
Conclusion
The multiplex Candida Pan/ auris LAMP assay provides a rapid, highly sensitive, and specific alternative to qPCR for C. auris screening, while preserving broad Candida detection in a single isothermal reaction. Its improved analytical and clinical sensitivity suggests strong potential for use in active surveillance and infection-control programs, particularly in settings where timely identification and containment of C. auris are critical.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Emergency Medicine, College of Medicine, Korea University Guro Hospital, Seoul, Republic of Korea.
- Department of Laboratory Medicine, Korea University Guro Hospital, Korea University College of Medicine, Seoul, Republic of Korea.
ORCID-Profile: Chae Seung Lim, Min-Chul Cho
Förderung
- Korea University Guro Hospital
- Korea Health Industry Development Institute
Zitieren
APA
Jang, W. S., Choe, Y. L., Yoon, S. Y., Lim, C. S., & Cho, M. C. (2026). Development of a Loop-Mediated Isothermal Amplification (LAMP) for the screening of Candida auris. PLOS One, 21(4), e0348003. https://doi.org/10.1371/journal.pone.0348003
Vancouver
Jang WS, Choe YL, Yoon SY, Lim CS, Cho MC. Development of a Loop-Mediated Isothermal Amplification (LAMP) for the screening of Candida auris. PLOS One. 2026;21(4):e0348003. doi:10.1371/journal.pone.0348003
BibTeX
@article{Jang2026,
author = {Jang, Woong Sik and Choe, Young Lan and Yoon, Soo Young and Lim, Chae Seung and Cho, Min-Chul},
title = {Development of a Loop-Mediated Isothermal Amplification (LAMP) for the screening of Candida auris},
journal = {PLOS One},
year = {2026},
volume = {21},
number = {4},
pages = {e0348003},
doi = {10.1371/journal.pone.0348003}
}
EUVIMED