Continuous remote monitoring of postoperative vital signs with the Biobeat patch: A multicenter prospective observational study
PLOS One · 20.04.2026 · 21 (4) · e0344662
DOI: 10.1371/journal.pone.0344662 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift PLOS One, online erschienen am 20.04.2026. Verfasst von Inès Kouidri und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesiology and Pain Medicine, Hôpital Foch, Suresnes, France. Hauptthemen laut MEDLINE: Vital Signs, Wearable Electronic Devices. Die Arbeit nennt 60 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
This study, conducted under standard clinical conditions, evaluated the Biobeat wearable sensor in postoperative patients monitored over multiple days without human intervention. The study aimed to assess technical performance (vital sign data collection and artifact frequency), reliability (agreement with nurse-measured values), the sensor’s ability to detect vital sign abnormalities, patient satisfaction, and incidence of adverse skin reactions.
Design
This multicenter prospective observational study included patients who underwent abdominal surgery with a planned postoperative ward stay between December 2020 and December 2022.
Methods
Alongside routine care–comprising (2–3 times daily nurse assessments of blood pressure (BP), pulse rate (PR), respiratory rate (RR), peripheral oxygen saturation (SpO2), and temperature)–participants wore a precordial Biobeat sensor. The sensor transmitted data wirelessly to a cloud-based repository via Wi-Fi without human intervention.
Results
Of 109 enrolled patients, 90 were included in the analysis. The median recording duration was 63.8 hours (interquartile range: 42.6–72.3 hours). Data loss occurred in 45% of the recording time, and pulse oximetry data were absent in 16% of the available measurements. Artifacts were infrequent, comprising 277 of 35,998 measurement points. Compared with nurse measurements, Biobeat showed small mean differences for most variables, except for RR and temperature, yet exhibited wide limits of agreement across all variables. Clarke Error Grid analysis revealed excellent concordance between Biobeat and nurse measurements for SpO2 and PR, with lower agreement for mean BP, RR, and temperature. The sensor’s continuous monitoring detected more vital sign abnormalities, including severe hypotension (mean BP < 60 mmHg), in 10.0% of patients compared to 3.3% with nurse measurements (P = 0.04). The patients reported high satisfaction and no adverse skin reactions were observed.
Conclusions
Significant data loss presents a substantial challenge for analysis, underscoring the critical need for improved data transmission and flow control measures in study protocols and clinical deployment. Registry ClinicalTrials.gov (NCT04585178, October 14, 2020).
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Anesthesiology and Pain Medicine, Hôpital Foch, Suresnes, France.
- Department of Epidemiology - Data - Biostatistics, Delegation of Clinical Research and Innovation, Hôpital Foch, Suresnes, France.
- Department of Anesthesiology, Groupe Hospitalier Diaconesses Croix Saint Simon, Paris, France.
- Department of Anesthesiology, Groupe Hospitalier Paris Saint-Joseph, Paris, France.
- Department of Information Systems, Hôpital Foch, Suresnes, France.
- Université Versailles-Saint-Quentin-en-Yvelines, Versailles, France.
ORCID-Profile: Morgan Le Guen, Marc Fischler
Zitieren
APA
Kouidri, I., Landais, C., Solis, A., Petit, D., Blondeau-Martin, A., Merlin, K., Trillat, B., Le Guen, M., & Fischler, M. (2026). Continuous remote monitoring of postoperative vital signs with the Biobeat patch: A multicenter prospective observational study. PLOS One, 21(4), e0344662. https://doi.org/10.1371/journal.pone.0344662
Vancouver
Kouidri I, Landais C, Solis A, Petit D, Blondeau-Martin A, Merlin K, et al. Continuous remote monitoring of postoperative vital signs with the Biobeat patch: A multicenter prospective observational study. PLOS One. 2026;21(4):e0344662. doi:10.1371/journal.pone.0344662
BibTeX
@article{Kouidri2026,
author = {Kouidri, Inès and Landais, Cécile and Solis, Audrey and Petit, Delphine and Blondeau-Martin, Aurélie and Merlin, Karine and Trillat, Bernard and Le Guen, Morgan and Fischler, Marc},
title = {Continuous remote monitoring of postoperative vital signs with the Biobeat patch: A multicenter prospective observational study},
journal = {PLOS One},
year = {2026},
volume = {21},
number = {4},
pages = {e0344662},
doi = {10.1371/journal.pone.0344662}
}
EUVIMED