Gait assessment using a 2D video-based motion analysis app in healthy subjects and subjects with lower limb amputation – A pilot study
PLOS One · 30.05.2025 · 20 (5) · e0324499
DOI: 10.1371/journal.pone.0324499 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS One, online erschienen am 30.05.2025. Verfasst von Frithjof Doerks und 4 weiteren Autor:innen. Beteiligte Einrichtung: Hannover Medical School, Department of Orthopaedic Surgery, DIAKOVERE Annastift, Laboratory for Biomechanics and Biomaterials, Hannover, Germany. Hauptthemen laut MEDLINE: Video Recording, Gait Analysis, Lower Extremity, Mobile Applications. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Although three-dimensional marker-based motion analysis is the gold standard for biomechanical investigations, it is time-consuming and cost-intensive. The conjunction of monocular video recordings with pose estimation algorithms addresses this gap. With the Orthelligent VISION app (OPED GmbH) a commercial and easy-to-use tool is now available for implementation in everyday clinical practice. The study investigates the accuracy of the 2D video-based system in measuring joint kinematics, expressed as range of motion, compared to an optoelectronic 3D motion analysis system as the gold standard. Materials and methods Its accuracy was determined by synchronously measuring ten healthy subjects with Orthelligent and the optoelectronic 3D motion analysis system Qualisys (Qualisys AB) during level walking and at different treadmill walking speeds (1 m/s; 1.4 m/s; 1.8 m/s). Range of motion (RoM) of lower limb joints and time-distance parameters were compared using Bland-Altman plots, t-tests, and correlations between systems. Kinematic outputs of two subjects with a lower limb amputation were also analyzed.
Results
The mean RoM deviation was smaller for the knee (3.8°) and hip joints (3.7°) than for the ankle joint (5.4°), but differed significantly between systems in most conditions. The correlation range was 0.36 ≤ r ≤ 0.83, with best results for 1 m/s treadmill walking (mean r = 0.71 across joints). While the accuracy was affected by high inter-subject variability, individual RoM changes from slow to fast walking did not differ between the systems. The kinematics of the prosthetic and sound leg of individuals with an amputation exhibited characteristic patterns in the video-based system, even though side differences were smaller compared to the optoelectronic measurement.
Conclusions
The rather high inter-subject variability would make future comparisons between individuals challenging. Nonetheless, the app shows potential for intra-subject progress monitoring.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Hannover Medical School, Department of Orthopaedic Surgery, DIAKOVERE Annastift, Laboratory for Biomechanics and Biomaterials, Hannover, Germany.
- Department for Medical Technology, Bremerhaven University of Applied Sciences, Bremerhaven, Germany.
ORCID-Profile: Michael Schwarze, Bastian Welke
Zitieren
APA
Doerks, F., Harms, F., Schwarze, M., Jakubowitz, E., & Welke, B. (2025). Gait assessment using a 2D video-based motion analysis app in healthy subjects and subjects with lower limb amputation – A pilot study. PLOS One, 20(5), e0324499. https://doi.org/10.1371/journal.pone.0324499
Vancouver
Doerks F, Harms F, Schwarze M, Jakubowitz E, Welke B. Gait assessment using a 2D video-based motion analysis app in healthy subjects and subjects with lower limb amputation – A pilot study. PLOS One. 2025;20(5):e0324499. doi:10.1371/journal.pone.0324499
BibTeX
@article{Doerks2025,
author = {Doerks, Frithjof and Harms, Fenna and Schwarze, Michael and Jakubowitz, Eike and Welke, Bastian},
title = {Gait assessment using a 2D video-based motion analysis app in healthy subjects and subjects with lower limb amputation – A pilot study},
journal = {PLOS One},
year = {2025},
volume = {20},
number = {5},
pages = {e0324499},
doi = {10.1371/journal.pone.0324499}
}
EUVIMED