Vergleichsstudie in der Zeitschrift PLOS ONE, 2024. Verfasst von Ji-Hoon Baek und 6 weiteren Autor:innen. Beteiligte Einrichtung: Joint & Arthritis Research, Department of Orthopaedic Surgery, Himchan Hospital, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Robotic Surgical Procedures, Osteotomy, Tibia. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study aimed to compare the clinical outcomes, mechanical axis, component positioning, leg length discrepancy (LLD), and polyethylene liner thickness between robotic-assisted total knee arthroplasty (TKA) and conventional TKA in patients with failed high tibial osteotomy (HTO).

Methods

A total of 30 patients (30 knees) with failed HTO who underwent TKA using a robot-assisted system between June 2020 and December 2023 were included in this study (robotic group). Additionally, 60 patients (60 knees) with failed HTO who underwent conventional TKA were included as controls (conventional group). Propensity score matching was performed using a 2:1 ratio between the matched participants. The mean follow-up period was 2.1 years in the robotic group and 2.2 years in the conventional group. Clinical evaluations were performed using the Knee Society Score (KSS) rating system. Mechanical axis, component coronal and sagittal positioning, and LLD were evaluated using postoperative radiographs. The thickness of the polyethylene liner was also determined. The mean error values and outliers were calculated and compared between the two groups to determine the accuracy of the mechanical axis, postoperative component positioning, and LLD.

Results

The postoperative KSSs in the robotic and conventional groups were not statistically different. The robotic group achieved better accuracy than the conventional group in terms of postoperative mean mechanical axis (1.7° vs. 2.4°, p < 0.05), femur coronal inclination (90.0° vs. 91.6°, p < 0.05), tibial coronal inclination (90.3° vs. 91.3°, p < 0.05), tibial sagittal inclination (90.5° vs. 91.4°, p < 0.05), and LLD (2.2 vs. 7.0 mm, p < 0.05). A significant difference in polyethylene liner thickness was observed between the two groups (p < 0.05).

Conclusions

Robotic-assisted TKA showed improved mechanical axis, higher accuracy of component positioning and polyethylene liner thickness, and reduced LLD compared with those of conventional TKA in patients with failed HTO. Further studies with a larger sample size and long-term follow-up are warranted to ascertain whether the accuracy of robotic-assisted TKA can translate into better clinical outcomes and patient satisfaction.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, KneeRobotic Surgical ProceduresOsteotomyTibiaHumansFemaleMaleMiddle AgedAgedTreatment OutcomeRetrospective StudiesKnee Joint

Autor:innen und Institutionen

  • Joint & Arthritis Research, Department of Orthopaedic Surgery, Himchan Hospital, Seoul, Republic of Korea.
  • Joint & Arthritis Research, Department of Neurosurgery, Himchan Hospital, Seoul, Republic of Korea.

ORCID-Profile: Chang Hyun Nam

Zitieren

APA

Baek, J. H., Lee, S. C., Lee, D. N., Heo, J., Kim, T., Ahn, H. S., & Nam, C. H. (2024). Better accuracy of robotic-assisted total knee arthroplasty compared to conventional technique in patients with failed high tibial osteotomy. PLOS ONE. https://doi.org/10.1371/journal.pone.0313391

Vancouver

Baek JH, Lee SC, Lee DN, Heo J, Kim T, Ahn HS, et al. Better accuracy of robotic-assisted total knee arthroplasty compared to conventional technique in patients with failed high tibial osteotomy. PLOS ONE. 2024. doi:10.1371/journal.pone.0313391

BibTeX
@article{Baek2024,
  author = {Baek, Ji-Hoon and Lee, Su Chan and Lee, Dong Nyoung and Heo, Juneyoung and Kim, Taehyeon and Ahn, Hye Sun and Nam, Chang Hyun},
  title = {Better accuracy of robotic-assisted total knee arthroplasty compared to conventional technique in patients with failed high tibial osteotomy},
  journal = {PLOS ONE},
  year = {2024},
  doi = {10.1371/journal.pone.0313391}
}
Kostenlos starten →