Investigation of the effectiveness of no-reference metric in image evaluation in nuclear medicine

Shigeaki Higashiyama iD, Yutaka Katayama, Atsushi Yoshida, Nahoko Inoue, Takashi Yamanaga, Takao Ichida, Yukio Miki, Joji Kawabe

PLOS ONE · 2024

DOI: 10.1371/journal.pone.0310305 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS ONE, 2024. Verfasst von Shigeaki Higashiyama und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nuclear Medicine, Graduate School of Medicine, Osaka Metropolitan University, Osaka, Japan. Hauptthemen laut MEDLINE: Nuclear Medicine, Phantoms, Imaging, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

In nuclear medicine, normalized mean square error (NMSE) is widely used for image quality evaluation and machine adjustment. However, evaluating clinical images in nuclear medicine using NMSE necessitates acquiring a reference image, which is time consuming and impractical. Therefore, it is necessary to explore no-reference metrics, such as perception-based image quality evaluator (PIQE) and natural image quality evaluator (NIQE), as alternatives for evaluating the quality of clinical images used in nuclear medicine.

Purpose

To examine whether no-reference metrics can be applied to image quality evaluations for clinical images in nuclear medicine.

Methods

Images of the Hoffman Brain Phantom containing 18F–fluoro-2-deoxy-D-glucose (FDG) were obtained using Biograph Vision (Siemens Co., Ltd). From the collected images, 14 images with varying pixel counts and acquisition times were created. Sixteen images were visually evaluated by five image experts and ranked accordingly. Image quality was assessed using NMSE, PIQE, and NIQE, and rankings were calculated based on these scores.

Results

The Spearman’s significance test revealed a strong correlation between image quality evaluations using PIQE and visual evaluations by specialists (p

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Nuclear MedicinePhantoms, ImagingImage Processing, Computer-AssistedHumansFluorodeoxyglucose F18BrainRadiopharmaceuticals

Substanzen

Fluorodeoxyglucose F18Radiopharmaceuticals

Autor:innen und Institutionen

  • Department of Nuclear Medicine, Graduate School of Medicine, Osaka Metropolitan University, Osaka, Japan.
  • Department of Radiology, Osaka Metropolitan University Hospital, Osaka, Japan.
  • Department of Diagnostic and Interventional Radiology, Graduate School of Medicine, Osaka Metropolitan University, Osaka, Japan.

ORCID-Profile: Shigeaki Higashiyama

Zitieren

APA

Higashiyama, S., Katayama, Y., Yoshida, A., Inoue, N., Yamanaga, T., Ichida, T., Miki, Y., & Kawabe, J. (2024). Investigation of the effectiveness of no-reference metric in image evaluation in nuclear medicine. PLOS ONE. https://doi.org/10.1371/journal.pone.0310305

Vancouver

Higashiyama S, Katayama Y, Yoshida A, Inoue N, Yamanaga T, Ichida T, et al. Investigation of the effectiveness of no-reference metric in image evaluation in nuclear medicine. PLOS ONE. 2024. doi:10.1371/journal.pone.0310305

BibTeX
@article{Higashiyama2024,
  author = {Higashiyama, Shigeaki and Katayama, Yutaka and Yoshida, Atsushi and Inoue, Nahoko and Yamanaga, Takashi and Ichida, Takao and Miki, Yukio and Kawabe, Joji},
  title = {Investigation of the effectiveness of no-reference metric in image evaluation in nuclear medicine},
  journal = {PLOS ONE},
  year = {2024},
  doi = {10.1371/journal.pone.0310305}
}
Kostenlos starten →