Development and performance evaluation of a solenoid valve assisted low-cost ventilator on gas exchange and respiratory mechanics in a porcine model

Daniel Shyu, Peter Bliss, Alexander Adams, Roy Joseph Cho iD

PLOS ONE · 16.05.2024 · 19 (5) · e0303443

DOI: 10.1371/journal.pone.0303443 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS ONE, online erschienen am 16.05.2024. Verfasst von Daniel Shyu und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medicine, University of Minnesota, Minneapolis, Minnesota, United States of America. Hauptthemen laut MEDLINE: Ventilators, Mechanical, COVID-19, Respiratory Mechanics, Pulmonary Gas Exchange. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Minnesota Emergency Response.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

During the COVID-19 pandemic, ventilator shortages necessitated the development of new, low-cost ventilator designs. The fundamental requirements of a ventilator include precise gas delivery, rapid adjustments, durability, and user-friendliness, often achieved through solenoid valves. However, few solenoid-valve assisted low-cost ventilator (LCV) designs have been published, and gas exchange evaluation during LCV testing is lacking. This study describes the development and performance evaluation of a solenoid-valve assisted low-cost ventilator (SV-LCV) in vitro and in vivo, focusing on gas exchange and respiratory mechanics.

Methods

The SV-LCV, a fully open ventilator device, was developed with comprehensive hardware and design documentation, utilizing solenoid valves for gas delivery regulation. Lung simulator testing calibrated tidal volumes at specified inspiratory and expiratory times, followed by in vivo testing in a porcine model to compare SV-LCV performance with a conventional ventilator.

Results

The SV-LCV closely matched the control ventilator’s respiratory profile and gas exchange across all test cycles. Lung simulator testing revealed direct effects of compliance and resistance changes on peak pressures and tidal volumes, with no significant changes in respiratory rate. In vivo testing demonstrated comparable gas exchange parameters between SV-LCV and conventional ventilator across all cycles. Specifically, in cycle 1, the SV-LCV showed arterial blood gas (ABG) results of pH 7.54, PCO2 34.5 mmHg, and PO2 91.7 mmHg, compared to the control ventilator’s ABG of pH 7.53, PCO2 37.1 mmHg, and PO2 134 mmHg. Cycle 2 exhibited ABG results of pH 7.53, PCO2 33.6 mmHg, and PO2 84.3 mmHg for SV-LCV, and pH 7.5, PCO2 34.2 mmHg, and PO2 93.5 mmHg for the control ventilator. Similarly, cycle 3 showed ABG results of pH 7.53, PCO2 32.1 mmHg, and PO2 127 mmHg for SV-LCV, and pH 7.5, PCO2 35.5 mmHg, and PO2 91.3 mmHg for the control ventilator.

Conclusion

The SV-LCV provides similar gas exchange and respiratory mechanic profiles compared to a conventional ventilator. With a streamlined design and performance akin to commercially available ventilators, the SV-LCV presents a viable, readily available, and reliable short-term solution for overcoming ventilator supply shortages during crises.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Ventilators, MechanicalCOVID-19Respiratory MechanicsPulmonary Gas ExchangeAnimalsSwineEquipment DesignRespiration, ArtificialSARS-CoV-2Tidal Volume

Autor:innen und Institutionen

  • Department of Medicine, University of Minnesota, Minneapolis, Minnesota, United States of America.
  • Independent Researcher, Philips Respironics, Minneapolis, Minnesota, United States of America.
  • Department of Respiratory Therapy, Hennepin County Medical Center, Minneapolis, Minnesota, United States of America.
  • Department of Medicine, Section of Interventional Pulmonology, University of Minnesota, Minneapolis, Minnesota, United States of America.

ORCID-Profile: Roy Joseph Cho

Förderung

  • University of Minnesota Emergency Response

Zitieren

APA

Shyu, D., Bliss, P., Adams, A., & Cho, R. J. (2024). Development and performance evaluation of a solenoid valve assisted low-cost ventilator on gas exchange and respiratory mechanics in a porcine model. PLOS ONE, 19(5), e0303443. https://doi.org/10.1371/journal.pone.0303443

Vancouver

Shyu D, Bliss P, Adams A, Cho RJ. Development and performance evaluation of a solenoid valve assisted low-cost ventilator on gas exchange and respiratory mechanics in a porcine model. PLOS ONE. 2024;19(5):e0303443. doi:10.1371/journal.pone.0303443

BibTeX
@article{Shyu2024,
  author = {Shyu, Daniel and Bliss, Peter and Adams, Alexander and Cho, Roy Joseph},
  title = {Development and performance evaluation of a solenoid valve assisted low-cost ventilator on gas exchange and respiratory mechanics in a porcine model},
  journal = {PLOS ONE},
  year = {2024},
  volume = {19},
  number = {5},
  pages = {e0303443},
  doi = {10.1371/journal.pone.0303443}
}
Kostenlos starten →