Vector analysis of corneal astigmatism in cataractous eyes based on IOLMaster 700 biometry
PLOS ONE · 2024
DOI: 10.1371/journal.pone.0300576 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS ONE, 2024. Verfasst von Achim Langenbucher und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Experimental Ophthalmology, Saarland University, Homburg/Saar, Germany. Hauptthemen laut MEDLINE: Astigmatism, Cataract Extraction, Corneal Diseases. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
The purpose of this study was to investigate the effect of the corneal back surface by comparing the keratometric astigmatism (K, derived from the corneal front surface) of a modern optical biometer against astigmatism of Total Keratometry (TK, derived from both corneal surfaces) in a large population with cataractous eyes. The results were then used to define linear prediction models to map K to TK.
Methods
From a large dataset containing bilateral biometric measurements (IOLMaster 700) in 9736 patients prior to cataract surgery, the total corneal astigmatism was decomposed into vectors for K, corneal back surface (BS), and TK. A multivariate prediction model (MV), simplified model with separation of vector components (SM) and a constant model (CM) were defined to map K to TK vector components.
Results
The K centroid (X/Y) showed some astigmatism with-the-rule (0.1981/-0.0211 dioptre (dpt)) whereas the TK centroid was located around zero (-0.0071/-0.0381 dpt against-the-rule) and the BS centroid showed systematic astigmatism against-the-rule (-0.2367/-0.0145 dpt). The respective TK–K centroid was located at -0.2052/-0.0302 dpt. The MV model showed the same performance (i.e. mean absolute residuum) as the SM did (0.1098 and 0.1099 dpt respectively) while the CM performed only slightly worse (0.1121 dpt mean absolute residuum).
Conclusion
In cases where tomographic data are unavailable statistical models could be used to consider the overall contribution of the back surface to the total corneal astigmatism. Since the performance of the CM is sufficiently close to that of MV and SM we recommend using the CM which can be directly considered e.g. as surgically induced astigmatism.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Experimental Ophthalmology, Saarland University, Homburg/Saar, Germany.
- Department of Ophthalmology, Johannes Kepler University Linz, Linz, Austria.
- School of Physical Sciences, The Open University, Milton Keynes, United Kingdom.
- Augen- und Laserklinik Castrop-Rauxel, Castrop-Rauxel, Germany.
- Dr. Rolf M. Schwiete Center for Limbal Stem Cell and Aniridia Research, Saarland University, Homburg/Saar, Germany.
- Department of Ophthalmology, Semmelweis-University, Budapest, Hungary.
ORCID-Profile: Achim Langenbucher
Zitieren
APA
Langenbucher, A., Wendelstein, J., Cayless, A., Hoffmann, P., & Szentmáry, N. (2024). Vector analysis of corneal astigmatism in cataractous eyes based on IOLMaster 700 biometry. PLOS ONE. https://doi.org/10.1371/journal.pone.0300576
Vancouver
Langenbucher A, Wendelstein J, Cayless A, Hoffmann P, Szentmáry N. Vector analysis of corneal astigmatism in cataractous eyes based on IOLMaster 700 biometry. PLOS ONE. 2024. doi:10.1371/journal.pone.0300576
BibTeX
@article{Langenbucher2024,
author = {Langenbucher, Achim and Wendelstein, Jascha and Cayless, Alan and Hoffmann, Peter and Szentmáry, Nóra},
title = {Vector analysis of corneal astigmatism in cataractous eyes based on IOLMaster 700 biometry},
journal = {PLOS ONE},
year = {2024},
doi = {10.1371/journal.pone.0300576}
}
EUVIMED