A scoping review on the clinical effectiveness of Trans-Impedance Matrix (TIM) measurements in detecting extracochlear electrodes and tip fold overs in Cochlear Ltd devices
PLOS ONE · 2024
DOI: 10.1371/journal.pone.0299597 ↗Kurz gefasst
Scoping Review in der Zeitschrift PLOS ONE, 2024. Verfasst von Muhammed Ayas und 7 weiteren Autor:innen. Beteiligte Einrichtung: College of Health Sciences, University of Sharjah, Sharjah, United Arab Emirates. Hauptthemen laut MEDLINE: Cochlear Implantation, Cochlear Implants. Die Arbeit wurde bisher 21-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Extrusion of electrodes outside the cochlea and tip fold overs may lead to suboptimal outcomes in cochlear implant (CI) recipients. Intraoperative measures such as Trans-Impedance Matrix (TIM) measurements may enable clinicians to identify electrode malposition and direct surgeons to correctly place the electrode array during surgery.
Objectives
To assess the current literature on the effectiveness of TIM measurements in identifying extracochlear electrodes and tip fold overs.
Methods
A scoping review of studies on TIM-based measurements were carried out using the Databases-Medline/PubMed, AMED, EMBASE, CINAHL and the Cochrane Library following PRISMA guidelines. Eleven full texts articles met the inclusion criteria. Only human studies pertaining to TIM as a tool used in CI were included in the review. Further, patient characteristics, electrode design, and TIM measurement outcomes were reported.
Results
TIM measurements were available for 550 implanted ears with the subjects age ranged between 9 months to 89 years. Abnormal TIM measurements were reported for 6.55% (36). Tip fold over was detected in 3.64% (20) of the cases, extracochlear electrodes in 1.45% (8), and 1.45% (8) were reported as buckling. Slim-modiolar electrode array designs were more common (54.71%) than pre-curved (23.34%) or lateral wall (21.95%) electrode array. Abnormal cochlear anatomy was reported for five ears (0.89%), with normal cochlear anatomy for all other patients.
Conclusion
TIM measurement is a promising tool for the intraoperative detection of electrode malposition. TIM measurement has a potential to replace intraoperative imaging in future. Though, TIM measurement is in its early stages of clinical utility, intuitive normative data sets coupled with standardised criteria for detection of abnormal electrode positioning would enhance its sensitivity.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- College of Health Sciences, University of Sharjah, Sharjah, United Arab Emirates.
- Emmeline Centre, Cambridge University Hospitals NHS Foundation Trust, Cambridge, United Kingdom.
- Cambridge Hearing Group, University of Cambridge, Cambridge, United Kingdom.
- Department of Ear, Nose and Throat Surgery, University Hospitals Birmingham NHS Foundation Trust, Birmingham, United Kingdom.
- Department of ENT, Cambridge University Hospitals NHS Foundation Trust, Cambridge, United Kingdom.
- Medical Library, University of Cambridge School of Clinical Medicine, Cambridge, United Kingdom.
ORCID-Profile: Muhammed Ayas, Jameel Muzaffar, Veronica Phillips
Zitieren
APA
Ayas, M., Muzaffar, J., Borsetto, D., Eitutis, S., Phillips, V., Tam, Y. C., Salorio-Corbetto, M., & Bance, M. L. (2024). A scoping review on the clinical effectiveness of Trans-Impedance Matrix (TIM) measurements in detecting extracochlear electrodes and tip fold overs in Cochlear Ltd devices. PLOS ONE. https://doi.org/10.1371/journal.pone.0299597
Vancouver
Ayas M, Muzaffar J, Borsetto D, Eitutis S, Phillips V, Tam YC, et al. A scoping review on the clinical effectiveness of Trans-Impedance Matrix (TIM) measurements in detecting extracochlear electrodes and tip fold overs in Cochlear Ltd devices. PLOS ONE. 2024. doi:10.1371/journal.pone.0299597
BibTeX
@article{Ayas2024,
author = {Ayas, Muhammed and Muzaffar, Jameel and Borsetto, Daniele and Eitutis, Susan and Phillips, Veronica and Tam, Yu Chuen and Salorio-Corbetto, Marina and Bance, Manohar L},
title = {A scoping review on the clinical effectiveness of Trans-Impedance Matrix (TIM) measurements in detecting extracochlear electrodes and tip fold overs in Cochlear Ltd devices},
journal = {PLOS ONE},
year = {2024},
doi = {10.1371/journal.pone.0299597}
}
EUVIMED