Cochlear coverage with lateral wall cochlear implant electrode arrays affects post-operative speech recognition
PLOS ONE · 2023
DOI: 10.1371/journal.pone.0287450 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS ONE, 2023. Verfasst von Tobias Weller und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Otorhinolaryngology, Hannover Medical School, Hannover, Germany. Hauptthemen laut MEDLINE: Cochlear Implants, Speech Perception, Cochlear Implantation. Die Arbeit wurde bisher 32-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Deutsche Forschungsgemeinschaft, Bundesministerium für Bildung und Forschung.
Abstract (Originalsprache)
Objectives
The goal was to investigate the relationship between the insertion angle/cochlear coverage of cochlear implant electrode arrays and post-operative speech recognition scores in a large cohort of patients implanted with lateral wall electrode arrays.
Methods
Pre- and post-operative cone beam computed tomography scans of 154 ears implanted with lateral wall electrode arrays were evaluated. Traces of lateral wall and electrode arrays were combined into a virtual reconstruction of the implanted cochlea. This reconstruction was used to measure insertion angles and proportional cochlear coverage. Word recognition scores and sentence recognition scores measured 12 months after implantation using electric-only stimulation were used to examine the relationship between cochlear coverage/insertion angle and implantation outcomes.
Results
Post-operative word recognition scores and the difference between post- and pre-operative word recognition scores were positively correlated with both cochlear coverage and insertion angle, however sentence recognition scores were not. A group-wise comparison of word recognition scores revealed that patients with cochlear coverage below 70% performed significantly worse than patients with coverage between 79%-82% (p = 0.003). Performance of patients with coverage above 82% was on average poorer than between 79%-82, although this finding was not statistically significant (p = 0.84). Dividing the cohort into groups based on insertion angle quadrants revealed that word recognition scores were highest above 450° insertion angle, sentence recognition scores were highest between 450° and 630° and the difference between pre- and post-operative word recognition scores was largest between 540° and 630°, however none of these differences reached statistical significance.
Conclusions
The results of this study show that cochlear coverage has an effect on post-operative word recognition abilities and the benefit patients receive from their implant. Generally, higher coverage led to better outcomes, however there were results indicating that insertion past 82% cochlear coverage may not provide an additional benefit for word recognition. These findings can be useful for choosing the optimal electrode array and thereby improving cochlear implantation outcomes on a patient-individual basis.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Otorhinolaryngology, Hannover Medical School, Hannover, Germany.
- German Hearing Center (DHZ), Hannover, Germany.
- Cluster of Excellence "Hearing4All", Oldenburg, Germany.
ORCID-Profile: Tobias Weller
Förderung
- Deutsche Forschungsgemeinschaft
- Bundesministerium für Bildung und Forschung
Zitieren
APA
Weller, T., Timm, M. E., Lenarz, T., & Büchner, A. (2023). Cochlear coverage with lateral wall cochlear implant electrode arrays affects post-operative speech recognition. PLOS ONE. https://doi.org/10.1371/journal.pone.0287450
Vancouver
Weller T, Timm ME, Lenarz T, Büchner A. Cochlear coverage with lateral wall cochlear implant electrode arrays affects post-operative speech recognition. PLOS ONE. 2023. doi:10.1371/journal.pone.0287450
BibTeX
@article{Weller2023,
author = {Weller, Tobias and Timm, Max Eike and Lenarz, Thomas and Büchner, Andreas},
title = {Cochlear coverage with lateral wall cochlear implant electrode arrays affects post-operative speech recognition},
journal = {PLOS ONE},
year = {2023},
doi = {10.1371/journal.pone.0287450}
}