Validity and intrarater reliability of a novel device for assessing Plantar flexor strength

Seth O'Neill iD, Alice Weeks, Jens Eg Nørgaard iD, Martin Gronbech Jorgensen

PLOS ONE · 2023

DOI: 10.1371/journal.pone.0282395 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS ONE, 2023. Verfasst von Seth O'Neill und 3 weiteren Autor:innen. Beteiligte Einrichtung: School of Healthcare, College of Life sciences, University of Leicester, Leicester, United Kingdom. Hauptthemen laut MEDLINE: Achilles Tendon, Tendinopathy. Die Arbeit wurde bisher 18-mal zitiert (Crossref) und nennt 51 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Plantar flexor weakness is an identified prospective factor for developing Achilles tendinopathy. Various authors have reported relationships between symptoms and weakness of this muscle group. Despite this relationship, many clinicians and researchers fail to examine Plantar flexor strength due to the cumbersome, stationary and expensive nature of an isokinetic dynamometer (IKD), known as the “Gold Standard”. This study examined the validity and reliability of a fast, easy and portable device for assessing plantarflexion.

Methods

Validity between the Cybex NORM® by Humac and the C-Station by Fysiometer was explored using Pearson correlation coefficient. Participants were randomly selected to start in the Cybex NORM® or the FysioMeter C-Station. Intra-rater reliability on the C-station was investigated by test-retest two days apart using Intraclass Correlation Coefficient (ICC). All testing involved isometric maximal force of the soleus muscle with the knee at 90 degrees flexion.

Results

40 healthy university students were recruited for the validity part, while 65 healthy university students were recruited for the reliability part of the study. The mean peak torque on the IKD was 198.55Nm (SD 94.45) versus 1443.88 (412.82)N on the C-Station. The results of the Pearson correlation revealed an r-value of r = 0.72 with a 95%CI 0.52–0.84. The test re-test reliability was calculated as an ICC of 0.91 with a (95%CI 0.86–0.94).

Conclusions

The C-Station by Fysiometer appears to provide valid measures and have excellent reliability for Plantar flexor isometric strength. It would appear suitable for both clinical and research work.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Achilles TendonTendinopathyHumansReproducibility of ResultsProspective StudiesMuscle StrengthMuscle Strength DynamometerMuscle, SkeletalIsometric Contraction

Autor:innen und Institutionen

  • School of Healthcare, College of Life sciences, University of Leicester, Leicester, United Kingdom.
  • Department of Geriatric Medicine, Aalborg University Hospital, Aalborg, Denmark.
  • Department of Clinical Medicine, Aalborg University, Aalborg, Denmark.

ORCID-Profile: Seth O'Neill, Jens Eg Nørgaard

Zitieren

APA

O'Neill, S., Weeks, A., Nørgaard, J. E., & Jorgensen, M. G. (2023). Validity and intrarater reliability of a novel device for assessing Plantar flexor strength. PLOS ONE. https://doi.org/10.1371/journal.pone.0282395

Vancouver

O'Neill S, Weeks A, Nørgaard JE, Jorgensen MG. Validity and intrarater reliability of a novel device for assessing Plantar flexor strength. PLOS ONE. 2023. doi:10.1371/journal.pone.0282395

BibTeX
@article{ONeill2023,
  author = {O'Neill, Seth and Weeks, Alice and Nørgaard, Jens Eg and Jorgensen, Martin Gronbech},
  title = {Validity and intrarater reliability of a novel device for assessing Plantar flexor strength},
  journal = {PLOS ONE},
  year = {2023},
  doi = {10.1371/journal.pone.0282395}
}
Kostenlos starten →