Hybrid calibration algorithm based on RKHS-PLSR for high-accuracy noninvasive optical monitoring of blood parameters
Optics Express · 18.05.2026 · 34 (10) · 17426
DOI: 10.1364/oe.589961 ↗Kurz gefasst
Fachartikel in der Zeitschrift Optics Express, online erschienen am 04.05.2026. Verfasst von Yuzhuo Miao und 8 weiteren Autor:innen. Hauptthemen laut MEDLINE: Algorithms. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 23 Literaturangaben. Gefördert von: National Key Research and Development Program of China, National Major Scientific Instruments and Equipments Development Project, Tianjin Metrology Technology Project.
Abstract (Originalsprache)
During the clinical extracorporeal life support process, noninvasive optical measurements of hematocrit (HCT) and blood oxygen saturation (SO 2 ) often suffer from limited accuracy and a need for frequent calibration, particularly for HCT, and cannot yet provide truly continuous, high-accuracy monitoring. To address these limitations, a hybrid calibration algorithm combining the reproducing kernel Hilbert space (RKHS) with partial least squares regression (PLSR) is proposed. The algorithm first constructs a physically interpretable baseline model via the modified Lambert–Beer law and then integrates an RKHS-based anomaly suppression mechanism to inversely correct outlier interference in the original optical signals. This framework ultimately leverages the PLSR algorithm to conduct multiparameter joint calibrations that couple multiwavelength light intensity features with physical prediction outcomes. This hybrid framework effectively reduces the recalibration frequency while increasing the detection accuracy. A custom-designed multiwavelength inline optical monitoring system was developed and validated using blood samples with varying HCT and SO 2 levels. The experimental results demonstrate that, compared with conventional PLSR, the RKHS-PLSR algorithm reduces the root mean square errors (RMSE) of the HCT and SO 2 measurements to 1.41% and 1.23%, respectively. These results demonstrate that this innovative fusion algorithm, which synergizes physical correction with data calibration, significantly enhances the accuracy and long-term stability of continuous optical monitoring within extracorporeal circuits, reduces the recalibration frequency, and improves clinical reliability.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Förderung
- National Key Research and Development Program of China
- National Major Scientific Instruments and Equipments Development Project
- Tianjin Metrology Technology Project
Zitieren
APA
Miao, Y., Liu, Y., Huang, Z., Kong, D., He, H., Zhang, H., Fan, H., Wu, T., & Jia, D. (2026). Hybrid calibration algorithm based on RKHS-PLSR for high-accuracy noninvasive optical monitoring of blood parameters. Optics Express, 34(10), 17426. https://doi.org/10.1364/oe.589961
Vancouver
Miao Y, Liu Y, Huang Z, Kong D, He H, Zhang H, et al. Hybrid calibration algorithm based on RKHS-PLSR for high-accuracy noninvasive optical monitoring of blood parameters. Optics Express. 2026;34(10):17426. doi:10.1364/oe.589961
BibTeX
@article{Miao2026,
author = {Miao, Yuzhuo and Liu, Yuhan and Huang, Zhanzhe and Kong, Deyu and He, Huicheng and Zhang, Hongxia and Fan, Haojun and Wu, Ting and Jia, Dagong},
title = {Hybrid calibration algorithm based on RKHS-PLSR for high-accuracy noninvasive optical monitoring of blood parameters},
journal = {Optics Express},
year = {2026},
volume = {34},
number = {10},
pages = {17426},
doi = {10.1364/oe.589961}
}
EUVIMED