Biomechanical finite element analysis of a femoral neck system fixation construct for femur neck fractures and clinical implications

Hoon-Sang Sohn, Se-Lin Jeong, Gu-Hee Jung

Journal of Musculoskeletal Trauma · 2025

DOI: 10.12671/jmt.2025.00108 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Musculoskeletal Trauma, 2025. Verfasst von Hoon-Sang Sohn und 2 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 3.0). Gefördert von: Gyeongsang National University.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

This study assessed the structural/mechanical stability of fixation constructs with a femoral neck system (FNS) via finite element analysis after simulating femoral neck fractures and explored the clinical implications.

Methods

We simulated subcapital, transcervical, basicervical, and vertical fracture models using a right femur (SAWBONES) and imported the implant model of FNS to Ansys (Ansys 19.0, Ansys Inc.) to place the implant in the optimal position. The distal end of the femur model was completely fixed and was abducted 7°. The force vector was set laterally at an angle of 3° and posteriorly at an angle of 15° in the vertical ground. The analysis was conducted using Ansys software with the von Mises stress (VMS) in megapascals (MPa).

Results

The maximum VMS of the fracture site was 67.01 MPa for a subcapital, 68.56 MPa for a transcervical, 344.54 MPa for a basicervical, and 130.59 MPa for a vertical model. The maximum VMS of FNS was 840.34 MPa for a subcapital, 637.37 MPa for a transcervical, 464.07 MPa for a basicervical, and 421.01 MPa for a vertical model. The stress distribution of basicervical and vertical fractures differed significantly, and the basicervical fracture had higher VMS at the bone, implant, and fracture sites.

Conclusions

FNS fixation should be performed with consideration the osseous anchorage in the femoral head, and this technique might be appropriate for vertical fractures. Regarding the VMS at the fracture site, FNS might be applied cautiously only to basicervical fractures with anatomical reduction without a gap or comminution. Level of evidence: IV.

Förderung

  • Gyeongsang National University

Zitieren

APA

Sohn, H., Jeong, S., & Jung, G. (2025). Biomechanical finite element analysis of a femoral neck system fixation construct for femur neck fractures and clinical implications. Journal of Musculoskeletal Trauma. https://doi.org/10.12671/jmt.2025.00108

Vancouver

Sohn H, Jeong S, Jung G. Biomechanical finite element analysis of a femoral neck system fixation construct for femur neck fractures and clinical implications. Journal of Musculoskeletal Trauma. 2025. doi:10.12671/jmt.2025.00108

BibTeX
@article{Sohn2025,
  author = {Sohn, Hoon-Sang and Jeong, Se-Lin and Jung, Gu-Hee},
  title = {Biomechanical finite element analysis of a femoral neck system fixation construct for femur neck fractures and clinical implications},
  journal = {Journal of Musculoskeletal Trauma},
  year = {2025},
  doi = {10.12671/jmt.2025.00108}
}
Kostenlos starten →