Implementation of automated personalised breast radiotherapy planning techniques with scripting in Raystation
The British Journal of Radiology · 2023
DOI: 10.1259/bjr.20220707 ↗Kurz gefasst
Fachartikel in der Zeitschrift The British Journal of Radiology, 2023. Verfasst von Ian Gleeson und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Cambridge University Hospitals NHS Foundation Trust, Hills Road, Cambridge, UK. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Cancer Research UK.
Abstract (Originalsprache)
Objective
Implement scripted automatic breast planning (AP) for breast techniques within Raystation.
Methods
Manual plans (MPs) were re-planned and compared with AP plans for whole breast (WB), partial breast (PB), hybrid volumetric modulated arc therapy simultaneous integrated boost (VMAT SIB) and VMAT nodal plans.
Results
WB AP plans took 7 min comparing well to MP. One WB AP failed a mandatory dose constraint. Small statistically significant differences showed improved coverage for AP at expense of slightly hotter plans, however absolute differences were small (mean differences < 1% or D 0.5cc
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Medical Physics, Cambridge University Hospitals NHS Foundation Trust, Hills Road, Cambridge, UK.
ORCID-Profile: Ian Gleeson
Förderung
- Cancer Research UK
Zitieren
APA
Gleeson, I., Bolger, N., Chun, H., Hutchinson, K., Klodowska, M., Mehrer, J., & Toomey, M. (2023). Implementation of automated personalised breast radiotherapy planning techniques with scripting in Raystation. The British Journal of Radiology. https://doi.org/10.1259/bjr.20220707
Vancouver
Gleeson I, Bolger N, Chun H, Hutchinson K, Klodowska M, Mehrer J, et al. Implementation of automated personalised breast radiotherapy planning techniques with scripting in Raystation. The British Journal of Radiology. 2023. doi:10.1259/bjr.20220707