EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Automatic consultation system for patients with cardiac implantable electronic devices undergoing magnetic resonance imaging

Donghoon Han, Ji Hyun Lee, Il-Young Oh, In-ho Chae, Ho-Young Lee, Youngjin Cho

International Journal of Arrhythmia · 2020

DOI: 10.1186/s42444-019-0010-3 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Arrhythmia, 2020. Verfasst von Donghoon Han und 5 weiteren Autor:innen. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Safety evaluation for patients with cardiac implantable electronic devices (CIEDs) undergoing magnetic resonance imaging (MRI) scanning is often overlooked. We developed an automatic consultation system (ACS) to improve the screening rate in these patients.

Methods

ACS was developed by the Hospital Information System Development Department of Seoul National University Bundang Hospital. It was designed to automatically request pre-MRI cardiac evaluation in patients with CIED when MRI orders are issued. The proportion of the patients without pre-MRI cardiologic evaluation was evaluated before and after the ACS application.

Results

From January 2016 to June 2018, a total of 157 patients with CIEDs [pacemaker 136 (86.6%), ICD or CRT-D 21 (13.4%), MR-conditional 117 (74.5%)] visited the MRI facility. Before the ACS application, 23 out of 84 patients (27.4%) did not have adequate pre-MRI cardiologic evaluation. Despite urgent request for pre-MRI cardiac evaluation, MRI examination was postponed or cancelled in 14 (60.8%) cases. After the ACS application, all 73 patients underwent proper cardiologic evaluation before their MRI examinations ( P < 0.001). The proportion of immediate request for pre-MRI evaluation at the moment of MRI order also improved with the ACS application (before ACS 57.1%, after ACS 100%, P < 0.001).

Conclusions

The newly developed ACS helped the patients with CIED receive MRI scan safely on the schedule, improving the quality of care in this population.

Zitieren

APA

Han, D., Lee, J. H., Oh, I., Chae, I., Lee, H., & Cho, Y. (2020). Automatic consultation system for patients with cardiac implantable electronic devices undergoing magnetic resonance imaging. International Journal of Arrhythmia. https://doi.org/10.1186/s42444-019-0010-3

Vancouver

Han D, Lee JH, Oh I, Chae I, Lee H, Cho Y. Automatic consultation system for patients with cardiac implantable electronic devices undergoing magnetic resonance imaging. International Journal of Arrhythmia. 2020. doi:10.1186/s42444-019-0010-3

BibTeX
@article{Han2020,
  author = {Han, Donghoon and Lee, Ji Hyun and Oh, Il-Young and Chae, In-ho and Lee, Ho-Young and Cho, Youngjin},
  title = {Automatic consultation system for patients with cardiac implantable electronic devices undergoing magnetic resonance imaging},
  journal = {International Journal of Arrhythmia},
  year = {2020},
  doi = {10.1186/s42444-019-0010-3}
}
Kostenlos starten →