High insertion torque versus regular insertion torque: early crestal bone changes on dental implants in relation to primary stability—a retrospective clinical study

Jörg-Martin Ruppin, Michael Stimmelmayr

International Journal of Implant Dentistry · 2024

DOI: 10.1186/s40729-024-00540-3 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Implant Dentistry, 2024. Verfasst von Jörg-Martin Ruppin und 1 weiteren Autor:innen. Beteiligte Einrichtung: , Penzberg, Germany. dr.ruppin@iz-ruppin.de. Hauptthemen laut MEDLINE: Torque, Dental Implants, Dental Implantation, Endosseous. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The aim of the presented retrospective study was to evaluate the early crestal bone changes around an implant type designed for high primary stability. A total number of 111 implants placed clinically were evaluated regarding insertion torque, bone density, implant stability quotient (ISQ) and early crestal bone loss from standardized digital radiographs. The implants were allocated in two groups: the „regular torque “ group contained all implants that achieved less than 50 Ncm as final insertion torque (n = 63) and the „high torque“ group contained the implants that achieved 50–80 Ncm (n = 48). To avoid possible damage either to the implant´s inner connection or to the bone by application of excessive force, a limit of 80 Ncm was set for all surgeries. All implants underwent submerged healing for three months. ISQ measurements and standardized digital radiographs were taken at day of insertion and at day of second stage surgery. The bone loss was measured on the mesial and distal aspect of the implant. The data evaluation showed the following results: Mean bone loss was 0.27 ± 0.30 mm for the high torque group and 0.24 ± 0.27 mm for the regular torque group. The difference was not statistically significant (p = 0.552). In the two groups, no complications nor implant loss occurred. For the evaluated implant type, there was no significant difference in crestal bone changes and complication rate between high and regular insertion torque in the early healing period. Graphical Abstract

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

TorqueDental ImplantsDental Implantation, EndosseousHumansRetrospective StudiesFemaleMaleMiddle AgedBone DensityAlveolar Bone LossAgedAdult

Stichwörter der Autor:innen

Bone densityBone lossImmediate loadingInsertion torquePrimary stabilityRFA analysis

Substanzen

Dental Implants

Autor:innen und Institutionen

  • , Penzberg, Germany. dr.ruppin@iz-ruppin.de.
  • Department of Prosthetic Dentistry, University Hospital, LMU Munich, Munich, Germany.
  • , Cham, Germany.

Zitieren

APA

Ruppin, J. M., & Stimmelmayr, M. (2024). High insertion torque versus regular insertion torque: early crestal bone changes on dental implants in relation to primary stability—a retrospective clinical study. International Journal of Implant Dentistry. https://doi.org/10.1186/s40729-024-00540-3

Vancouver

Ruppin JM, Stimmelmayr M. High insertion torque versus regular insertion torque: early crestal bone changes on dental implants in relation to primary stability—a retrospective clinical study. International Journal of Implant Dentistry. 2024. doi:10.1186/s40729-024-00540-3

BibTeX
@article{Ruppin2024,
  author = {Ruppin, Jörg-Martin and Stimmelmayr, Michael},
  title = {High insertion torque versus regular insertion torque: early crestal bone changes on dental implants in relation to primary stability—a retrospective clinical study},
  journal = {International Journal of Implant Dentistry},
  year = {2024},
  doi = {10.1186/s40729-024-00540-3}
}
Kostenlos starten →