A novel Z-shaped anti-rotation rod for atlantoaxial dislocation reduction: finite element analysis

Mandi Cai, Rencai Ma, Junlin Chen, Xinzhao Huang, Yixing Zhang, Zhuohang Xie, Xiaobao Zou, Xiangyang Ma

Journal of Orthopaedic Surgery and Research · 2025

DOI: 10.1186/s13018-025-05723-1 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Orthopaedic Surgery and Research, 2025. Verfasst von Mandi Cai und 7 weiteren Autor:innen. Beteiligte Einrichtung: The First School of Clinical Medicine, Southern Medical University, No.1838 North of Guangzhou Road, Guangzhou, 510515, People's Republic of China. Hauptthemen laut MEDLINE: Finite Element Analysis, Atlanto-Axial Joint, Joint Dislocations, Pedicle Screws. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Natural Science Foundation of Guangdong Province, National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

C1-C2 pedicle screw-rod fixation (PSR) is widely used for atlantoaxial dislocations. However, its limited reduction capacity in refractory cases necessitates additional release surgeries, increasing operative risks including prolonged surgical time and expanded tissue damage. We developed a novel Z-shaped anti-rotation rod to improve reduction capability, but its biomechanical performance requires evaluation.

Methods

A nonlinear atlantoaxial instability three-dimensional (3D) C0-C3 finite element model was constructed using computed tomography images from a 25-year-old healthy male without a history of cervical spine diseases. Based on this model, two C1-C2 fixation configurations were simulated: conventional pedicle screw-rod (PS-CR) and pedicle screw-Z-shaped rod (PS-ZR). Reduction forces were measured and compared. Range of motion (ROM), stress distribution and peak stress values of the implants were recorded and compared under six loading conditions including flexion, extension, lateral bending, and axial rotation.

Results

Both configurations achieved a greater than 98% reduction in the C1-C2 segmental ROM, with similar compensatory motions in adjacent segments. The reduction force of PS-ZR showed significant advantages (2–8 mm range), achieving a maximum reduction force of 88.544 N, which is 1.67 to 3.68 times that of PS-CR. The PS-ZR system experiences greater stress compared to the PS-CR system, escalating with Z-rod height. Regarding stress distribution and peak values of rods, the maximum stress on the PS-CR system was mainly concentrated at the connection between the rod and the screw nut while the maximum stress on the PS-ZR system was concentrated at the transition part of the “Z” shape.

Conclusions

Both PS-CR and PS-ZR configurations provide reliable and comparable stability. Compared to the PS-CR configuration, the PS-ZR configuration provides superior reduction force and stability, potentially reducing the need for additional release surgery and surgical time. This novel design has significant clinical implications for improving fixation techniques.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Finite Element AnalysisAtlanto-Axial JointJoint DislocationsPedicle ScrewsHumansMaleAdultRange of Motion, ArticularBiomechanical PhenomenaRotationTomography, X-Ray ComputedImaging, Three-Dimensional

Stichwörter der Autor:innen

Anti-rotationAtlantoaxial dislocationBiomechanicsFinite element analysisInternal fixatorsZ-shaped rod

Autor:innen und Institutionen

  • The First School of Clinical Medicine, Southern Medical University, No.1838 North of Guangzhou Road, Guangzhou, 510515, People's Republic of China.
  • Department of Orthopedics, General Hospital of Southern Theatre Command of PLA, 111 Liuhua Road, Guangzhou, 510010, People's Republic of China.
  • Department of Orthopedics, General Hospital of Southern Theatre Command of PLA, 111 Liuhua Road, Guangzhou, 510010, People's Republic of China. zouxb6478@126.com.
  • The First School of Clinical Medicine, Southern Medical University, No.1838 North of Guangzhou Road, Guangzhou, 510515, People's Republic of China. maxy1001@126.com.
  • Department of Orthopedics, General Hospital of Southern Theatre Command of PLA, 111 Liuhua Road, Guangzhou, 510010, People's Republic of China. maxy1001@126.com.

Förderung

  • Natural Science Foundation of Guangdong Province
  • National Natural Science Foundation of China

Zitieren

APA

Cai, M., Ma, R., Chen, J., Huang, X., Zhang, Y., Xie, Z., Zou, X., & Ma, X. (2025). A novel Z-shaped anti-rotation rod for atlantoaxial dislocation reduction: finite element analysis. Journal of Orthopaedic Surgery and Research. https://doi.org/10.1186/s13018-025-05723-1

Vancouver

Cai M, Ma R, Chen J, Huang X, Zhang Y, Xie Z, et al. A novel Z-shaped anti-rotation rod for atlantoaxial dislocation reduction: finite element analysis. Journal of Orthopaedic Surgery and Research. 2025. doi:10.1186/s13018-025-05723-1

BibTeX
@article{Cai2025,
  author = {Cai, Mandi and Ma, Rencai and Chen, Junlin and Huang, Xinzhao and Zhang, Yixing and Xie, Zhuohang and Zou, Xiaobao and Ma, Xiangyang},
  title = {A novel Z-shaped anti-rotation rod for atlantoaxial dislocation reduction: finite element analysis},
  journal = {Journal of Orthopaedic Surgery and Research},
  year = {2025},
  doi = {10.1186/s13018-025-05723-1}
}
Kostenlos starten →