EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Feasibility and accuracy of continuous intraabdominal pressure monitoring with a capsular device in human pilot trial

Chien-Hung Liao, David A Spain, Chih-Chi Chen, Chi-Tung Cheng, Wei-Cheng Lin, Dong-Ru Ho, Heng-Fu Lin, Fausto Catena

World Journal of Emergency Surgery · 2025

DOI: 10.1186/s13017-024-00569-0 ↗

Kurz gefasst

Fachartikel in der Zeitschrift World Journal of Emergency Surgery, 2025. Verfasst von Chien-Hung Liao und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Trauma and Emergency Surgery, Chang Gung Memorial Hospital, Chang Gung University, Taoyuan, Taiwan. Hauptthemen laut MEDLINE: Intra-Abdominal Hypertension. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Science and Technology Council, Chang Gung Medical Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Intrabdominal pressure (IAP) is an important parameter. Elevated IAP can reduce visceral perfusion, lead to intraabdominal hypertension, and result in life-threatening abdominal compartment syndrome. While ingestible capsular devices have been used for various abdominal diagnoses, their application in continuous IAP monitoring remains unproven.

Method

We conducted a prospective clinical trial to evaluate the feasibility of IAP measurement using a digital capsule PressureDOT, an ingestible capsule equipped with wireless transmission capability and a pressure sensor, then compared its reliability with conventional intravesical method. Patients undergoing laparoscopic or robotic surgeries were recruited. During surgery, we created pneumoperitoneum by inflating CO2 into the peritoneal cavity and IAP was simultaneously monitored using both the ingestible capsules and intravesical measurements from Foley catheter. We assessed the feasibility of signal transmission and the accuracy of pressure measurements.

Results

Six patients were enrolled in this pilot study. No adverse events were reported, and the average first-intake time was within 24 h. All capsules were successfully expelled, with an average excretion time of 81 h. In the summarized data, the mean IAPdot is 0.6 mmHg lower than the IAPivp, with a standard deviation of 1.68 mmHg. However, capsule measurements showed excellent correlation with intravesical IAP measurements, with an intraclass correlation coefficient of 0.916 (95% CI: 0.8821–0.9320).

Conclusion

Our study demonstrates the feasibility and safety of using digital capsules for continuous IAP monitoring, providing the agreement between IAP measurements from digital capsules and conventional intravesical measurement within a near-normal pressure.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Intra-Abdominal HypertensionHumansPilot ProjectsMaleProspective StudiesFemaleMiddle AgedFeasibility StudiesMonitoring, PhysiologicPressureAgedAdultReproducibility of ResultsAbdominal Cavity

Stichwörter der Autor:innen

Abdominal compartment syndromeCapsular sensorDigital healthIntraabdominal pressure

Autor:innen und Institutionen

  • Department of Trauma and Emergency Surgery, Chang Gung Memorial Hospital, Chang Gung University, Taoyuan, Taiwan.
  • Department of Surgery, Stanford University, Stanford, CA, USA.
  • Department of Physical Medicine and Rehabilitation, Chang Gung Memorial Hospital, Chang Gung University, Taoyuan, Taiwan.
  • Department of Electrical Engineering, Chang Gung University, Taoyuan, Taiwan.
  • Department of Surgery, Division of Urology, Chang Gung Memorial Hospital, Chang Gung University, Puzi City, Chiayi, Taiwan. redoxdrh@gmail.com.
  • School of Medicine, National Tsing Hua University, Hsinchu, Taiwan. redoxdrh@gmail.com.

Förderung

  • National Science and Technology Council
  • Chang Gung Medical Foundation

Zitieren

APA

Liao, C. H., Spain, D. A., Chen, C. C., Cheng, C. T., Lin, W. C., Ho, D. R., Lin, H. F., & Catena, F. (2025). Feasibility and accuracy of continuous intraabdominal pressure monitoring with a capsular device in human pilot trial. World Journal of Emergency Surgery. https://doi.org/10.1186/s13017-024-00569-0

Vancouver

Liao CH, Spain DA, Chen CC, Cheng CT, Lin WC, Ho DR, et al. Feasibility and accuracy of continuous intraabdominal pressure monitoring with a capsular device in human pilot trial. World Journal of Emergency Surgery. 2025. doi:10.1186/s13017-024-00569-0

BibTeX
@article{Liao2025,
  author = {Liao, Chien-Hung and Spain, David A and Chen, Chih-Chi and Cheng, Chi-Tung and Lin, Wei-Cheng and Ho, Dong-Ru and Lin, Heng-Fu and Catena, Fausto},
  title = {Feasibility and accuracy of continuous intraabdominal pressure monitoring with a capsular device in human pilot trial},
  journal = {World Journal of Emergency Surgery},
  year = {2025},
  doi = {10.1186/s13017-024-00569-0}
}
Kostenlos starten →