Use of segmented CT transmission map to avoid metal artifacts in PET images by a PET-CT device

Siroos Mirzaei, Michel Guerchaft, Christopher Bonnier, Peter Knoll, Michel Doat, Peter Braeutigam

BMC Nuclear Medicine · 2005

DOI: 10.1186/1471-2385-5-3 ↗

Kurz gefasst

Fachartikel in der Zeitschrift BMC Nuclear Medicine, 2005. Verfasst von Siroos Mirzaei und 5 weiteren Autor:innen. Beteiligte Einrichtung: Centre National PET, Clinique Ste, Thérèse, Luxembourg. mirzaei@gmx.at. Die Arbeit wurde bisher 30-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Attenuation correction is generally used to PET images to achieve count rate values independent from tissue densities. The goal of this study was to provide a qualitative comparison of attenuation corrected PET images produced by a PET-CT device (CT, 120 kV, 40 mAs, FOV 600 mm) with and without segmentation of transmission data (ACseg+ and ACseg-respectively).

Methods

The reconstructed images were compared to attenuation corrected images obtained with a high-energy transmission source (Cs-137 - 662 keV).Thirty oncologic patients were studied using CT and 137Cs for attenuation correction. All image data were acquired using the Gemini PET-CT scanner (Philips Medical Systems). It is an open PET-CT system that consists of the MX8000 multislice CT and the Allegro PET scanner arranged in a separable configuration. Images with ACseg+ and ACseg- were analyzed simultaneously in coronal, sagittal and transaxial planes. Two nuclear medicine physicians reviewed the image sets.

Results

The image quality in the area of metal implants was better with ACseg+ than ACseg-, without metal induced artifacts generally observed in CT corrected images. Further the images with ACseg+ were qualitatively comparable to those obtained with 137Cs attenuation correction.

Conclusions

In case of metal implants, PET studies corrected by CT should preferably use the ACseg+ method to avoid the image artifacts.

Autor:innen und Institutionen

  • Centre National PET, Clinique Ste, Thérèse, Luxembourg. mirzaei@gmx.at.

Zitieren

APA

Mirzaei, S., Guerchaft, M., Bonnier, C., Knoll, P., Doat, M., & Braeutigam, P. (2005). Use of segmented CT transmission map to avoid metal artifacts in PET images by a PET-CT device. BMC Nuclear Medicine. https://doi.org/10.1186/1471-2385-5-3

Vancouver

Mirzaei S, Guerchaft M, Bonnier C, Knoll P, Doat M, Braeutigam P. Use of segmented CT transmission map to avoid metal artifacts in PET images by a PET-CT device. BMC Nuclear Medicine. 2005. doi:10.1186/1471-2385-5-3

BibTeX
@article{Mirzaei2005,
  author = {Mirzaei, Siroos and Guerchaft, Michel and Bonnier, Christopher and Knoll, Peter and Doat, Michel and Braeutigam, Peter},
  title = {Use of segmented CT transmission map to avoid metal artifacts in PET images by a PET-CT device},
  journal = {BMC Nuclear Medicine},
  year = {2005},
  doi = {10.1186/1471-2385-5-3}
}
Kostenlos starten →