Analysis of Screw/Pedicle-Width Ratio and Accuracy in Navigated Versus 3D-Controlled Fluoroscopy-Guided Pedicle Screw Placement
Global Spine Journal · 2026-01 · 16 (1) · 278-286
DOI: 10.1177/21925682251343523 ↗Kurz gefasst
Fachartikel in der Zeitschrift Global Spine Journal, online erschienen am 19.05.2025. Verfasst von Jula Gierse und 5 weiteren Autor:innen. Beteiligte Einrichtung: BG Klinik Ludwigshafen, Department for Orthopedics and Trauma Surgery at Heidelberg University, Ludwigshafen, Germany. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Study Design Retrospective cohort study.
Objectives
Due to the close anatomic relationship between the pedicle and neurovascular structures, avoiding pedicle perforations is crucial in pedicle screw placement. Still, the use of larger screws has biomechanical advantages. Intraoperative 2D and 3D imaging, and navigation guidance can be used to combine both goals. The aim of this study was to compare the screw diameter/pedicle width ratio (SPR) and the screw placement accuracy for 3D controlled fluoroscopy-guided pedicle screw placement vs computer navigated transpedicular screw placement in the thoracic and lumbar spine.
Methods
200 cases of thoracic and lumbar pedicle screw placement, of which 100 cases were performed using intraoperative computer navigation and 100 cases were performed using fluoroscopy-guidance were retrospectively registered. In the fluoroscopy group, intraoperative 3D scans were performed to confirm implant position and allow for potential intraoperative revision. In addition to accuracy and SPR, demographics, dose reports, and procedure times were analyzed.
Results
Overall, 716 fluoroscopy-guided screws were compared with 740 screws placed using navigation. Screw accuracy (83.2% vs 90.5%; P = .001) and SPR (0.85 ± 0.17 vs 0.88 ± 0.21; P < .001) were significantly higher using navigation compared to fluoroscopic guidance. Furthermore, dose area product (28,545 ± 17,693 vs 20,638 ± 15,856 mGycm 2 ; P < .001), fluoroscopy time (223.6 ± 93.6 vs 92.3 ± 39.7 seconds; P < .001), and procedure time (154.0 ± 81.0 vs 119.7 ± 48.7 min; P = .004) were significantly lower using navigation.
Conclusions
Intraoperative navigation does not only increase the accuracy of pedicle screw placement, but also allows for the placement of larger screws relative to the pedicle width, which may have biomechanical advantages. Notably, contrary to other studies, the use of navigation did not increase patient radiation exposure or procedure time compared to fluoroscopic guidance.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- BG Klinik Ludwigshafen, Department for Orthopedics and Trauma Surgery at Heidelberg University, Ludwigshafen, Germany.
- Heidelberg University, Heidelberg, Germany.
- Mannheim Institute for Intelligent Systems in Medicine, Heidelberg University, Mannheim, Germany.
ORCID-Profile: Jula Gierse, Sven Y Vetter, Eric Mandelka
Zitieren
APA
Gierse, J., Zimmermann, F., Grützner, P. A., Stallkamp, J., Vetter, S. Y., & Mandelka, E. (2026). Analysis of Screw/Pedicle-Width Ratio and Accuracy in Navigated Versus 3D-Controlled Fluoroscopy-Guided Pedicle Screw Placement. Global Spine Journal, 16(1), 278-286. https://doi.org/10.1177/21925682251343523
Vancouver
Gierse J, Zimmermann F, Grützner PA, Stallkamp J, Vetter SY, Mandelka E. Analysis of Screw/Pedicle-Width Ratio and Accuracy in Navigated Versus 3D-Controlled Fluoroscopy-Guided Pedicle Screw Placement. Global Spine Journal. 2026;16(1):278-286. doi:10.1177/21925682251343523
BibTeX
@article{Gierse2026,
author = {Gierse, Jula and Zimmermann, Felix and Grützner, Paul A and Stallkamp, Jan and Vetter, Sven Y and Mandelka, Eric},
title = {Analysis of Screw/Pedicle-Width Ratio and Accuracy in Navigated Versus 3D-Controlled Fluoroscopy-Guided Pedicle Screw Placement},
journal = {Global Spine Journal},
year = {2026},
volume = {16},
number = {1},
pages = {278--286},
doi = {10.1177/21925682251343523}
}
EUVIMED