Bone density measurements adjacent to acetabular cups in total hip arthroplasty using dual-energy CT: an in vivo reliability and agreement study

Bo Mussmann iD, Poul Erik Andersen, Trine Torfing, Søren Overgaard

Acta Radiologica Open · 2018-09 · 7 (9) · 205846011879653

DOI: 10.1177/2058460118796539 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Acta Radiologica Open, online erschienen am 30.08.2018. Verfasst von Bo Mussmann und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Odense University Hospital, Odense, Denmark. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Measuring bone mineral density (BMD) around acetabular prosthetic components with computed tomography (CT) is challenged by the complex anatomy and metal artifacts. Three-dimensional (3D) segmentation is required for the analysis, but it is usually not practically applicable on current CT workstations Purpose To test the between-scan agreement and reliability of custom segmentation software for BMD measurements adjacent to cemented and uncemented acetabular cups in dual-energy CT (DECT). Material and Methods Twenty-four male patients with total hip arthroplasty were scanned and rescanned using 130-keV virtual monochromatic DECT images. Hemispherical regions of interest were defined slice-by-slice and BMD was calculated around the acetabular cup using custom segmentation software.

Results

In the uncemented cup, the mean BMD was 153 mg/cm 3 with a between-scan difference of 10 mg/cm 3 ( P < 0.0001). In the cemented cup, the mean BMD was 186 mg/cm 3 with a between-scan difference of 6 mg/cm 3 ( P = 0.15). In both uncemented and cemented cups the intraclass correlation coefficient between repeated measurements was >0.95 and narrow Bland–Altman Limits of Agreement.

Conclusion

BMD can be measured with high absolute between-scan agreement and good reliability adjacent to acetabular cemented and uncemented cups using DECT and segmentation software.

Schlagwörter

Stichwörter der Autor:innen

Segmentationbone losscomputed tomography bone densitydual energy computed tomographyhip loosening cup

Autor:innen und Institutionen

  • Department of Radiology, Odense University Hospital, Odense, Denmark.
  • Department of Clinical Research, University of Southern Denmark, Odense, Denmark.
  • Department of Orthopedic Surgery and Traumatology, Odense University Hospital, Odense, Denmark.

ORCID-Profile: Bo Mussmann

Zitieren

APA

Mussmann, B., Andersen, P. E., Torfing, T., & Overgaard, S. (2018). Bone density measurements adjacent to acetabular cups in total hip arthroplasty using dual-energy CT: an in vivo reliability and agreement study. Acta Radiologica Open, 7(9), 205846011879653. https://doi.org/10.1177/2058460118796539

Vancouver

Mussmann B, Andersen PE, Torfing T, Overgaard S. Bone density measurements adjacent to acetabular cups in total hip arthroplasty using dual-energy CT: an in vivo reliability and agreement study. Acta Radiologica Open. 2018;7(9):205846011879653. doi:10.1177/2058460118796539

BibTeX
@article{Mussmann2018,
  author = {Mussmann, Bo and Andersen, Poul Erik and Torfing, Trine and Overgaard, Søren},
  title = {Bone density measurements adjacent to acetabular cups in total hip arthroplasty using dual-energy CT: an in vivo reliability and agreement study},
  journal = {Acta Radiologica Open},
  year = {2018},
  volume = {7},
  number = {9},
  pages = {205846011879653},
  doi = {10.1177/2058460118796539}
}
Kostenlos starten →